Abstract

A new electrochemical sandwich immunoassay for the ultrasensitive detection of human MUC1 cancer biomarker using protein G-functionalized magnetic beads (MBs) and graphite-based screen-printed electrodes (SPEs) was developed. Magnetic beads were employed as the platforms for the immobilization and immunoreaction process. A pair of primary and secondary antibodies was used to capture the MUC1 protein. After labeling with a third antibody conjugated with horseradish peroxidase (HRP), the resulting conjugate was trapped at the surface of the graphite-based SPEs and MUC1 determination was carried out by differential pulse voltammetry (DPV) at 0.4 V upon H2O2 addition using acetaminophen (APAP) as the redox mediator. A linear relationship was obtained for the detection of human MUC1 over a range of 0–25 ppb with the lowest detection limit of 1.34 ppb when HRP was applied as a label. Preliminary experiments were performed using disposable electrochemical sensors in order to optimize some parameters (i.e., incubation times, concentrations, and blocking agent).

1. Introduction

MUCl mucin is an integral membrane glycoprotein expressed by most if not all “wet” epithelia, such as on bladder, breast, gastric, pancreas, ovary, and respiratory tract epithelium, with a molecular weight varying between 250 and 1000 kD [1].

In normal secretory epithelial cells, MUC1 is expressed at the apical plasma membrane. However, following malignant transformation, MUC1 may be expressed at high levels on the entire membrane surface as well as in the cytoplasm. In addition to the altered cellular localization, changes in glycosylation may occur during malignant transformation [2, 3]. Although the numbers of glycosidic links are increased, the carbohydrate chains are generally shorter [4]. The reduced glycosylation reveals protein epitopes, making MUC1 more accessible to specific antibodies. Thus, overexpression and altered glycosylation allow MUC1 to be used as a cancer marker.

Widespread use of tumor markers in healthcare will ultimately depend upon the detection of many tumor markers with high selectivity and sensitivity. This goal has not been obtained with conventional methods which have poor precision or are experiencing difficulties in realizing automatization [5]. Recently, electrochemical immunosensors have received considerable attention for the detection of tumor markers due to their high sensitivity and easy miniaturization [6, 7].

Most immunoassay techniques are based on the separation of free and bound immunospecies. In the electrochemical immunosensors, one of the immunoagents (antigen or antibody) is immobilized on a solid carrier, capturing/binding the analyte to form an immunocomplex. As an ideal platform candidate, magnetic beads (MBs) allow for quick and efficient purification and concentration from crude samples, thereby eliminating the need for most of the pretreatment steps, as well as the matrix effect from the samples. Additionally, MBs can provide a low detection limit since MBs have large surface area to immobilize biomolecules. More importantly, MBs allow a simple manipulation, the immunocomplex being first formed on the beads surface and then, in a second step, immobilized on the surface of the electrode [8], effectively reducing in this way the complexity and time required for sensing application [915].

Quantification of bound immunoagents is conducted by using labels covalently bound to the immunoagent with specific properties suitable for detection. As a result, the final stage of immunoassay is in fact the quantitative detection of the label. Enzymes are the most frequently used labels linked to the antigen or to the antibody to visualize the binding event [16]. However, direct electrical communication between the redox sites of enzymes and the electrode usually is not possible because the redox sites of the enzymes are located deep inside the protein shell [17]. Therefore, redox mediators are required to facilitate the electron transfer between enzyme and electrode. Indeed, in order to achieve high sensitivity, the amount of mediator, as well as that of the enzyme, needs to be sufficiently high. Horseradish peroxidase (HRP), an enzyme which catalyzes the oxidation of a wide variety of organic and inorganic substrates in the presence of H2O2, was also used [18].

One sandwich immunoassay (three-layer format) was employed to detect MUC1 with horseradish peroxidase-conjugated Ab3, as a label. The determination of MUC1 was established by DPV upon H2O2 additions to record the reduction current response of acetaminophen as the redox mediator in the case of HRP-labeled immunoassay. These immunosensors displayed high sensitivity and a low detection limit. To our knowledge, the use of MBs and graphite SPEs for the construction of an immunosensor in a sandwich format for the detection of MUC1 has not been reported so far, so it may have potential clinical application in the future.

2. Materials and Methods

2.1. Reagents and Apparatus

Dynabeads protein G-coated magnetic beads (superparamagnetic beads with the diameter of 2.8 μm) were purchased from Invitrogen. MUC1 monoclonal mouse antibody (Ab1), MUC1 recombinant protein (Ag), and MUC1 polyclonal rabbit antibody (Ab2) were provided by Novus Biological. Polyclonal goat anti-rabbit IgG labeled with HRP (HRP-Ab3) was obtained from DakoCytomation. Bovine serum albumin (BSA), sodium chloride, potassium chloride, magnesium chloride, and polyoxyethylene sorbitan monolaurate (Tween 20) were purchased from Sigma-Aldrich. Disodium hydrogen phosphate, sodium dihydrogen phosphate, potassium ferrocyanide (K4[Fe(CN)6]), potassium ferricyanide (K3[Fe(CN)6]), hydrogen peroxide, and tris(hydroxymethyl)aminomethane (TRIS) were obtained from Merck. All the solutions were prepared using Milli-Q grade water.

The buffers used in the experiments are PBS (pH 7.4): phosphate buffered saline, pH 7.4; PBS-T (pH 7.4): phosphate buffered saline, pH 7.4 with 0.005% w/w Tween 20; PBS (pH 7.2): phosphate buffered saline, pH 7.2; PBS-T (pH 7.2): phosphate buffered saline, pH 7.2 with 0.005% w/w Tween 20; TRIS: 50 mM TRIS buffer, 10 mM NaCl, pH 8; TRIS-T: TRIS 20 mM, 150 mM NaCl, 0.1% BSA with 0.005% w/w Tween 20, pH 7.4.

Electrochemical experiments were performed with an Autolab PGSTAT 30 digital potentiostat/galvanostat with a GPES 4.9 software package (Eco Chemie, the Netherlands). Graphite-based screen-printed electrodes were produced by a DEK 248 screen printing machine at laboratory or were provided by DropSens (Spain). They are disposable three-electrode cells consisting of a graphite working electrode, a silver pseudo-reference electrode, and a graphite counter electrode. All the measurements were carried out at room temperature (25°C).

The electrochemical impedance spectroscopy (EIS) measurements were carried out in the presence of 10 mM [Fe(CN)6]3/4− redox probe in PBS (pH 7.4). The impedance was measured in a frequency range from 100 mHz to 100 KHz at a dc potential of +0.13 V, the formal potential of the [Fe(CN)6]3/4− redox probe. Experimental spectra, presented in the form of complex plane diagrams (i.e., Nyquist plots), were fitted with proper equivalent circuits using the facilities of the FRA2 software 4.9 (Eco Chemie). Differential pulse parameters are modulation time, 0.05 s; interval time, 0.15 s; step potential, 5 mV; modulation amplitude, 70 mV; potential scan, from +0.2 to +0.8 V.

2.2. Preparation of the Immunosensor
2.2.1. Immobilization of Ab1 and Blocking Agent onto the MBs

For the fabrication of the sandwich immunosensor, illustrated in Figure 1, 100 μL of the MBs from stock solution was placed in a vial and separated from the suspension media with a magnetic concentrator (MPC-S, Invitrogen). Subsequently, the MBs were washed three times with PBS-T buffer (pH 7.4). After removing the solution, MBs were suspended in 200 μL of 50 ppm Ab1 solution prepared in PBS buffer (pH 7.4) and incubated for two hours in the rotating stand at the room temperature.

To minimize the effect of nonspecific binding on detection, the Ab1 functionalized MBs were washed three times with PBS-T buffer (pH 7.2) solution and then the free sites on the surface of MBs were blocked by using 5% BSA as blocking agent were prepared in PBS buffer. MBs were then washed three times with 500 μL of PBS-T buffer (pH 7.4), suspended in 500 μL of PBS buffer (pH 7.4), and stored at 4°C.

2.2.2. Binding of the MUC1 Protein

50 μL of the Ab1-BSA functionalized MBs suspension was mixed, after the removal of the supernatant, with 500 μL of MUC1 protein solution in TRIS of different concentrations, between 0 and 25 ppb, for 90 minutes of incubation time at room temperature. Unbound antigens in excess were removed by washing with TRIS-T buffer (pH 7.4).

2.2.3. Reaction with Ab2 and HRP Conjugated Ab3

The Ab1-BSA-Ag modified MBs were mixed with 200 μL of 5 ppm Ab2 solution prepared in PBS buffer (pH 7.4) for 90 min. Finally, after rinsing with TRIS-T buffer (pH 7.4), the MBs were incubated with 50 μL of a solution containing 1.5 mg/L of the HRP-Ab3 in TRIS-T buffer. After 30 min, MBs were washed three times and suspended in 50 μL of the same buffer.

3. Results and Discussion

The stepwise procedure of the sandwich immunosensor development is shown in Figure 1. In a typical experiment, Ab1 was immobilized on the surface of modified protein G magnetic beads (a), and then free sites were blocked with 5% blocking agent solution (b) followed by the reaction with MUC 1 antigen (c) and the reaction with the second antibody Ab2 (d).

After introducing Ab3 labeled with HRP (e) on MBs, DPV measurements were recorded 5 minutes after the deposition of 10 μL modified beads onto graphite-based SPEs and the covering of the electrode’s surface with substrate solution (f). The substrate solution is represented by 40 μL of H2O2 4 mM + 10 μL of PBS containing 2 mM of acetaminophen.

Preliminary experiments were performed using modified MB on disposable screen-printed electrodes in order to optimize some analytical parameters in terms of incubation time, the type of blocking agent, and antibodies concentration.

3.1. Optimization of Immobilization Conditions
3.1.1. Incubation Time and Concentration

The incubation time and concentrations required for antibodies immobilization and the antigen-antibody immunoreaction are important parameters that determine the performance of an immunosensor. To optimize the time required for the complete immunoreaction to take place, modified MBs were incubated in 5 ppb of MUC1 protein solution for various time periods for Ab1, MUC1 Ag, and Ab2. DPV measurements were then performed using these incubated MBs on graphite-based SPEs. Only functionalized MBs incubated for 120 min for Ab1, 90 min for MUC1 Ag, and 90 min for Ab2 (Figure 2(a)) exhibited a stable response indicating the complete interaction between them.

After optimizing the incubation times used in different steps in the construction of the immunosensor, the concentration of Ab1 and Ab2 had to be optimized as well (data not shown). In order to optimize the primary antibody concentration, the MBs were functionalized with 10 ppm, 50 ppm, and 100 ppm primary antibodies. An increase of the signal was observed for the concentration of the Ab1 of 50 ppm, respect to 10 ppm. For concentrations higher than 50 ppm, considerable variation of signal was not observed, indicating that after this concentration a steady state is achieved. The concentration of 50 ppm primary solution was therefore chosen for further experiments.

For the optimization of Ab2 concentration, 5 ppm, 10 ppm and 20 ppm of antibody solution were used. Between the signals obtained for 5 and 10 ppm of Ab2 concentration, a significant difference was not found. For concentrations higher than 10 ppm, the secondary antibody binds probably directly on the magnetic beads, leading to a nonspecific adsorption; therefore, 5 ppm was assigned as the optimized value for further analysis.

3.1.2. Blocking Agent

The type and incubation time of the blocking agent on the immunosensor response were also investigated (Figure 2(b)). Milk powder (casein) and BSA were used as blocking agents with different incubation times. The Ab1 functionalized MBs were incubated with both milk powder 5% and BSA 5% in PBS buffer (pH 7.2) for two hours and overnight, respectively, and the experiments were carried out as the above described procedure. Overnight incubation time with BSA 5% showed the best results.

3.2. Impedance Characterization of Immunosensor

Compared with other electrochemical methods, the impedance technique has the advantage that the system is investigated under stationary conditions as opposed to the wide potential window used in CVs [20]. The features of HRP-MBs-based immunosensor were investigated by EIS in 10 mM [Fe(CN)6]3/4− prepared in PBS buffer (pH 7.4) step by step (Figure 3). It could be seen that the diameter of the semicircle in the impedance spectrum of bare graphite SPEs (Figure 3, curve a) increased after the immobilization of Ab1 and BSA (Figure 3, curve b). A change in impedance was observed after incubation with MUC1 antigen and Ab2 (Figure 3, curve c). The reasonable explanation is the electrostatic repulsion between the proteins and [Fe(CN)6]3/4− anion, which indicates that the antibodies and MUC1 antigen have been successfully immobilized on MBs and obstructed electron transfer. When HRP-Ab3 was introduced, the interfacial resistance increased more, demonstrating that the sensing interface is constructed successfully (Figure 3, curve d).

3.3. Analytical Calibration

The sandwich immunoassay was used for quantitative analysis of MUC1 concentrations. The electrochemical measurements were carried out under optimum conditions using graphite-based SPE and the results are shown in Figure 4. First, 10 μL of modified MBs was deposited on the working electrode surface and the stocking buffer was carefully removed with the aid of a micropipette. Then, the graphite-based SPE was covered with 40 μL of 4 mM H2O2 prepared in PBS and 10 μL of PBS containing 2 mM of acetaminophen was added. After 5 min, DPV measurements were recorded for different concentrations of MUC1 protein: 0, 2, 3, 4, 5, 7.5, 10, 15, 20, and 25 ppb. The limit of detection (LOD) was defined as the lowest concentration of biomarker producing a peak current 3 times higher than the standard deviation of the peak current in the absence of MUC1 under identical conditions. A sigmoid curve for the current variation was observed over the range 0–25 ppb of MUC1 with LOD 1.34 ppb (Figure 4, inset A). Inset A shows calibration plot of the DPV signal against the MUC1 protein concentration with two linear domains.

3.4. MUC1 Analysis in Human Serum Samples

Direct determination of serum MUC1 level is very important in the early detection of breast and ovarian cancer. The MUC1 contents in unknown real human serum samples were quantified with the designed immunosensor. All experimental results are given in Table 1.

The samples were obtained from patients suffering from breast or ovarian cancer, filtered, diluted 1 : 200, and assayed as in the procedure previously described. The standard addition method was used in order to quantify the amount of the tumor markers in the samples. The limit of detection in serum was calculated interpolating the mean current of the blank plus three times its standard deviation in the linear equation.

Experimental results revealed that the proposed immunosensor showed good sensitivity with wider linear range.

4. Conclusions

In summary, a magnetic beads-based electrochemical immunosensor is described as a device for the quantitative detection of human MUC1 antigen. During incubation, primary antibodies were bound to the functionalized MBs via their Fc regions and successively attached to the antigen and secondary antibody. This paper provides new perspectives on using HRP as the biocatalyst in the detection of MUC1 cancer biomarker in the presence of hydrogen peroxide on GSP electrodes. The present method could be applied for the detection of MUC1 in real samples of ovarian and breast cancer.

Conflict of Interests

The authors declare that there is no conflict of interests regarding the publication of this paper.

Acknowledgment

The authors are grateful for the financial support from the Romanian National Authority for Scientific Research, CNCS-UEFISCDI for the Project PN-II-ID-PCE-2011-3-0355.