Research Article

Evaluation of the BD GeneOhm MRSA and VanR Assays as a Rapid Screening Tool for Detection of Methicillin-Resistant Staphylococcus aureus and Vancomycin-Resistant Enterococci in a Tertiary Hospital in Saudi Arabia

Table 3

Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of the BD GeneOhm Assay MRSA real-time PCR using nasal and groin swabs.

Value % (95% CI)Type of sample
Nasal swabs (728)Groin swabs (1539)

Sensitivity197.2% (0.83–0.99)100% (0.86–1)
Specificity299.4% (0.98–0.99)98.7% (0.97–0.99)
PPV389.7% (0.74–0.96)61.5% (0.47–0.74)
NPV499.9% (0.99–0.99)100% (0.99–1)

1The proportion of actual positives by PCR which are correctly identified as such.
2The proportion of actual negatives by PCR which are correctly identified as such.
3The proportion of positive test results with PCR who are correctly diagnosed.
4The proportion of negative test results with PCR who are correctly diagnosed.