Research Article

Auditory Brainstem Responses to Bone-Conducted Brief Tones in Young Children with Conductive or Sensorineural Hearing Loss

Table 2

Comparison of BC-ABR category to AC behavioural threshold data for 2000 Hz and 500 Hz.

BC-ABR category (dBnHL)

Normal35–60>60
2000 Hz
104 ears
Behavioural outcome category (dBHL)Normal28 (26.9%) 0 (0%)0 (0%)
30–651 (1.0%)25 (24.0%)3 (2.9%)
>650 (0%)1 (1.0%)46 (44.2%)

Normal25–45>50
500 Hz
47 ears
Behavioural outcome category (dBHL)Normal36 (76.6%) 0 (0%)0 (0%)
30–551 (2.1%)5 (10.6%)0 (0%)
>550 (0%)1 (2.1%)4 (8.5%)

Percent of total ears in parentheses.
BC ABR “normal” levels: 30 and 20 dBnHL for 2000 and 500 Hz.
AC (and BC) behavioural “normal” levels are ≤25 dBHL.
Most behavioural threshold data are AC (2000 Hz: 89/104; 500 Hz: 33/47). See text for details.