Research Article

Contralateral Ear Occlusion for Improving the Reliability of Otoacoustic Emission Screening Tests

Table 1

TEOAE data from six subjects comparing OAE levels with contralateral ear occluded versus open. values of paired Student’s -test results and significance are indicated.

SubjectTEOAE (dB) contra ear pluggedTEOAE (dB) contra ear open Difference (dB) valueSignificance

Noise level <25 dB SPL
PAP14.2114.110.10.42NO
ALL9.429.30.120.629NO
GLU13.113.4−0.30.471NO
LAR10.3510.050.30.46NO
HAR5.084.860.210.15NO
SKL13.213.180.0160.6109NO

Noise level c. 45 dB SPL
PAP14.5913.620.960.003YES
ALL9.6 8.351.250.044YES
GLU13.2212.280.930.112NO
LAR9.128.720.40.093NO
HAR5.755.60.130.604NO
SKL12.1211.650.460.0004YES

Noise level 55 dB SPL
PAP13.9812.881.1<0.0001YES
ALL8.557.71.050.0085YES
GLU13.5511.91.65<0.0001YES
LAR8.787.90.880.151NO
HAR5.755.00.210.021YES
SKL12.2811.920.3660.0197YES

Babble level 55 dB SPL
PAP13.6312.621.020.0113YES
ALL9.478.411.0570.0036YES
GLU13.5512.151.40.0002YES
LAR9.438.331.10.009YES
HAR5.955.380.560.035YES
SKL13.3512.950.40.0015YES