A New Bottle Design Decreases Hypoxemic Episodes during Feeding in Preterm Infants

Table 2

Oxygen desaturation events, expressed as median and (interquartile range).

Standard bottleNew bottle designP *

Percentage of feeding time SpO2 <90%8% (3–13)5% (2–11)<0.004
Percentage of feeding time SpO2 90% −94%13% (6–21) 8% (2–18)<0.0007
Number of desaturation events per infant10 (1–19)4 (1–8)<0.001
Time with SpO2 <90% (s)46 s (8.3–150 )30 s (6–96)<0.001
Mean SpO2 during feeding94 (91–96)96 (93–98)<0.0008

*Wilcoxon signed-0 rank test.