Research Article
Reproducibility of BiliCare™ Transcutaneus Bilirrubin Meter in Mexican Newborns
Table 4
BiliCare™ interapparatus assay.
| | Device 1 | Device 2 | Device 3 | Device 4 |
| Device 1 | ICC2A1 | 1 | 0.980 | 0.932 | 0.982 | Pearson | 1 | 0.973 | 0.945 | 0.964 |
| Device 2 | ICC2A1 | 0.980 | 1 | 0.957 | 0.983 | Pearson | 0.973 | 1 | 0.943 | 0.973 |
| Device 3 | ICC2A1 | 0.932 | 0.957 | 1 | 0.955 | Pearson | 0.945 | 0.943 | 1 | 0.973 |
| Device 4 | ICC2A1 | 0.982 | 0.983 | 0.955 | 1 | Pearson | 0.964 | 0.973 | 0.973 | 1 |
|
|