Research Article

Human Myocardial Protein Pattern Reveals Cardiac Diseases

Figure 2

Non-scaled schematic view of an LC-FT-ICR MS system with a quadrupole mass selector and a hexapole collision cell. In the electrospray ion source the formed ions are captured after they pass the electrospray capillary in an ion funnel, which increases the sensitivity of the system by roughly up to an order of magnitude. Ions are transferred through two ion funnels into a hexapole ion guide, where they can also be trapped. Ions are selected in the quadrupole mass selector and can undergo collision induced dissociation in the hexapole collision chamber which is at a relatively high pressure. The ICR cell is in the magnetic center of a 9.4T superconducting magnet. The vacuum system, not shown in the figure, consists of pumping stages down to the range of 10−10 mbar in the ultra high vacuum chamber of the ICR cell. The numbers shown are approximate pressures in different pumping stages.
342659.fig.002