Journals
Publish with us
Publishing partnerships
About us
Blog
International Journal of Polymer Science
Journal overview
For authors
For reviewers
For editors
Table of Contents
Special Issues
International Journal of Polymer Science
/
2012
/
Article
/
Tab 1
/
Research Article
Long-Term Biostability of Pet Vascular Prostheses
Table 1
Main characteristics of the explanted grafts. Reference number is the reference number given in the first clinical description of the series. (MDV: Microvel Double Velour; CDV: Cooley Double Velour; RL: remeshing line; GL: guide line).
Explant reference
Prosthetic type
Year of implantation
Year of explantation
Lifetime (months)
Reason for explantation
Location of implantation
Prosthetic segment affected
Ruptured area
A
CDV
1980
1993
156
Bilateral false-aneurysm
Axillo bi-femoral
limb
RL
B
CDV
1981
1995
168
Multiplefalse-aneurysms
Axillo bi-femoral
Entire prosthesis
RL, GL
C
CDV
1982
1999
204
False-aneurysmand dilatation
Aorto bi-femoral
limb
GL
D
CDV
1979
1997
204
False-aneurysm
Axillo bi-femoral
limb
RL, GL
E
CDV
1978
1998
240
False-aneurysm
Aorto bi-femoral
body
RL
F
MDV
1978
1998
240
False-aneurysm
Aorto bi-femoral
limb
GL