Research Article

Sequential Spectrophotometric Method for the Simultaneous Determination of Amlodipine, Valsartan, and Hydrochlorothiazide in Coformulated Tablets

Table 4

Assay validation sheet of the proposed spectrophotometric method for the simultaneous determination of AML, VAL, and HCT.

ParameterProposed method
AMLVALHCT

Accuracy (mean ± RSD)
Precision
 Repeatabilitya0.5350.6510.867
 Intermediate precisionb1.0611.2251.002
Robustnessc0.7040.6330.807
Linearity
 Slope0.01620.02580.0108
 Intercept0.00100.00510.0030
 Correlation coefficient ( )0.99980.99980.9998
 Range (μg/mL)4–324–446–20

The intraday ( ) average of three concentrations (12, 16, 20  g/mL) for AML, VAL, and HCT repeated three times within the day.
The interday ( ) average of three concentrations (12, 16, 20  g/mL) for AML, VAL, and HCT repeated three times in three days.
Robustness ( ) average of three concentrations (12, 16, 20  g/mL) for AML, VAL, and HCT analyzed using 75 and 70% methanol.