ISRN Analytical Chemistry
Volume 2014 (2014), Article ID 926213, 10 pages
http://dx.doi.org/10.1155/2014/926213
Research Article

Determination of Phenol and Chlorophenols at Single-Wall Carbon Nanotubes/Poly(3,4-ethylenedioxythiophene) Modified Glassy Carbon Electrode Using Flow Injection Amperometry

1Department of Chemistry, Addis Ababa University, P.O. Box 1176, Addis Ababa, Ethiopia
2Department of Chemistry and Chemical Technology, National University of Lesotho, P.O. Box 180, Mailroom 127, Roma, Lesotho

Received 28 February 2014; Accepted 19 March 2014; Published 16 April 2014

Academic Editors: P. Campíns-Falcó, F. Kandemirli, and J. J. Santana-Rodríguez

Copyright © 2014 Negussie Negash et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Phenol and chlorophenols were investigated using single-wall carbon nanotubes (SWCNT) and poly(3,4-ethylenedioxythiophene) (PEDOT) composite modified glassy carbon electrode (SWCNT/PEDOT/GCE) as a detector in flow injection system. Optimization of experimental variables such as the detection potential, flow rate, and pH of the carrier solution (0.1 M sodium acetate) for the determination of phenol (P), 4-chlorophenol (CP), 2,4-dichlorophenol (DCP), 2,4,6-trichlorophenol (TCP), and pentachlorophenol (PCP) were performed. Under these conditions, analytical parameters were calculated from the calibration curve of measured amperometric responses as a function of concentrations of phenol and chlorophenols. The designed electrode exhibited very good analytical performance. The designed electrode was tested with 20 repetitive injections of each analyte and showed good operational stability. The analytical performance of the SWCNT/PEDOT/GCE electrode under flow through conditions was tested and was found to be impressive. The electrode showed a wider dynamic range for the detection of phenol and chlorophenols with low limits of detection compared with other enzymatic and nonenzymatic sensors. These results suggest that the method is quite useful for the analysis and monitoring of phenols and chlorophenols.