ISRN Oncology
Volume 2014 (2014), Article ID 125020, 5 pages
http://dx.doi.org/10.1155/2014/125020
Clinical Study

Treatment Planning Methods in High Dose Rate Interstitial Brachytherapy of Carcinoma Cervix: A Dosimetric and Radiobiological Analysis

1Department of Radiation Oncology, HCG Bangalore Institute of Oncology, RRMR Extension KH Road, Bangalore 560027, India
2Photonics, Nuclear and Medical Physics Division, VIT University, Vellore 632014, India
3Department of Radiation Oncology, Kidwai Memorial Institute of Oncology, Hosur Road, Bangalore 560029, India

Received 8 October 2013; Accepted 19 November 2013; Published 23 January 2014

Academic Editors: R.-J. Bensadoun and G. E. Kim

Copyright © 2014 Surega Anbumani et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Treatment planning is a trial and error process that determines optimal dwell times, dose distribution, and loading pattern for high dose rate brachytherapy. Planning systems offer a number of dose calculation methods to either normalize or optimize the radiation dose. Each method has its own characteristics for achieving therapeutic dose to mitigate cancer growth without harming contiguous normal tissues. Our aim is to propose the best suited method for planning interstitial brachytherapy. 40 cervical cancer patients were randomly selected and 5 planning methods were iterated. Graphical optimization was compared with implant geometry and dose point normalization/optimization techniques using dosimetrical and radiobiological plan quality indices retrospectively. Mean tumor control probability was similar in all the methods with no statistical significance. Mean normal tissue complication probability for bladder and rectum is 0.3252 and 0.3126 (), respectively, in graphical optimized plans compared to other methods. There was no significant correlation found between Conformity Index and tumor control probability when the plans were ranked according to Pearson product moment method (). Graphical optimization can result in maximum sparing of normal tissues.

1. Introduction

Cervical cancer is a major cancer burden which constitutes number one among Indian women with relative survival rate of 48.7% [1]. EBRT followed with brachytherapy (BT) is the standard of care and an integral part of local control of the disease. Cervix carcinoma has been treated with HDR BT for more than 30 years. It has advantages in terms of local recurrence, mortality, and late complications for the clinical stages I, II and III, similar to those of Low Dose Rate therapy [2]. The present standard of care using concomitant chemotherapy and radiotherapy has resulted in 80–90% local control rates for early stages [35]. But a decline of about 67–75% is noted for advanced stages because of the local failure due to inadequate dose coverage [68].

Interstitial brachytherapy (ISBT) is best suited for the patients with anatomy not allowing for standard intracavitary application and wherein the disease could not be encompassed in the standard ICBT application. It can improve the dose coverage with various normalization and optimization techniques. Improved planning strategies and dose optimization can reduce normal tissue complications rate without compromising local control of the disease [9]. Clinical investigations and dosimetric comparisons in the evaluation of interstitial implants are scarce and heterogeneous. Available data are from limited number of patients. Thus, the potential advantage of ISBT in gynecological malignancies could not be clearly demonstrated [10]. Interstitial implants, two dimensionally planned with semiorthogonal films, were radiobiologically evaluated using a figure of merit (FOM) and biologically effective dose by Supe et al. [11]. Different optimization techniques used in computed tomography based planning were dosimetrically compared in 10 patients by Shwetha et al., in which dose normalization methods were not considered [12]. The authors from Tata Memorial Hospital, India, have dosimetrically compared the anatomy based inverse planning with the graphical optimization for five prostate cases [13]. The studies conducted were on lesser number of patients and evaluated with mean and standard deviation values which could not establish a statistically significant result. None of these studies comprehensively evaluated dosimetrical and radiobiological indices.

HDR brachytherapy has the advantage of calculating dose distributions by altering radioactive source dwell times and dwell positions using sophisticated planning methods. Dose point normalization (DP_N) is a procedure where dose can be normalized to prescription dose along the target outline using dose points defined. The prescribed dose is optimized in the earlier defined dose points oriented to the target’s surface in the dose point optimization method. Reference isodose line can be normalized to a 5 mm box surface of the needle implant in geometrical normalization. In Geometrical Optimization procedure, dwell weight of each catheter position is primarily influenced by active dwell positions on other catheters, therefore filling in the spaces to cover desired volume. Dose optimization algorithms in general are said to be effective in optimizing the dose distributions by reducing the high dose regions inside the target volume. But the clinical advantage of each optimization technique should be evaluated.

Geometrical optimization was the previously followed ISBT planning technique in our institution. Radiation dose can be normalized using dose point and geometrical normalization techniques. The graphical optimization was compared with DP_N, DP_O, G_N, and G_O. In this study we have evaluated the ISBT planning methods for 40 patient cases with seven dosimetric parameters [14]. In addition to dosimetric analysis, the cervix implants were assessed with TCP and bladder and rectum NTCP estimates formulated by Kehwar et al. based on biologically effective equivalent uniform dose concept. Thus, the optimal ISBT planning method to improve local control of cervical cancer can be suggested.

2. Methods

40 cervical cancer patients (stage IIB/stage IIIB) treated with High Dose Rate Microselectron unit (V2) were selected for this retrospective study. Interstitial needles were inserted using vaginal obturator (VO) in the template (either Nucletron MUPIT or Syed Neblett) under general anesthesia through transperineal route. First a guide needle is inserted into the anterior lip of cervix and the other 5 needles were inserted into the groove of the VO. The total number of needles used ranges from 20 to 26 depending upon disease extension. During the implant procedure, utmost care was taken in order not to perforate bladder and rectum with the needles. Patients with needles in situ were CT scanned (slice thickness 5 mm) and the images were converted into DICOM RT files. The CT images were acquired into Plato Sun Rise Treatment Planning System Version v 3.5.1. In the virtual simulation module of Plato Version VSS v1.6 (Nucletron, Elekta AB, The Netherlands), target volume and critical structures such as bladder and rectum were delineated by the physicians (clinical oncologists). Catheters were reconstructed in the brachytherapy planning module BPS v 14.3. In our institution, dose point based optimization followed by graphical optimization is the planning method in practice. Dose points for normalizing the prescribed dose (6 Gy per fraction) were created at an interspace distance of 5 mm on the target volume. Only those dwell positions of the source with step size 2.5 mm which encompass the target volume alone were activated manually. Then the dose points were normalized to prescription dose and then optimized. Further optimization by manual dragging of isodose lines was done with the graphical optimization tool. It is used to customize the shape of an isodose distribution along the target volume. It allows the manipulation of isodose curves and modifies the individual dwell weights or dwell times by dragging the isodose lines visualized in the axial CT sections. This is a trial and error process in which the dwell weight/dwell time are adjusted to achieve an optimal target coverage with less bladder (D2 = 5 Gy) and rectum doses (D2 = 4 Gy). Hence it requires a very high expertise of the planner. Other available planning methods in Plato BPS which need to be validated and compared [15].

CT scans of 40 patients previously treated with Gr_O were used to construct hypothetical treatment plans using DP_N, DP_O, G_N, and G_O. Hence, all the 40 patients planned with graphical optimization technique (Gr_O) were kept as their own internal control in this study. Dose volume histograms (CTV, Bladder, and Rectum) for each patient with five plans were generated that resulted in a total of 200 (40 × 5) plans and a sum of 600 DVHs (200 × 3). The dosimetric parameters such as TVDref, TV1.5Dref, TV2Dref, D2cc (Bl), and D2cc (R) were extracted from the DVHs. From these parameters, the dosimetric quality predictors such as Coverage Index (CI), External Volume Index (EI), Dose Homogeneity Index (DHI), Overdose Volume Index (ODI), and Dose Nonuniformity Index (DNR) defined by Meertens et al. were computed (Table 1): CI is the fraction of the target volume that receives a dose equal to or greater than reference dose. EI is defined as the ratio of the volume of normal tissue that receives a dose equal to or greater than reference dose to target volume whereas DHI is the ratio of target volume receiving a dose in the range of 1.0 to 1.5 times of reference dose [16, 17]. ODI is the ratio of the target volume which receives a dose equal to or more than 2.0 times of the reference dose. DNR is calculated from the ratio of the target volume which receives a dose equal to or greater than 1.5 times of reference dose [18]. For an ideal implant, CI is 1, EI is 0, DHI is 1, ODI is 0, and DNR is 0.

tab1
Table 1: Dosimetrical and radiobiological data for the ISBT planning methods.

Radiobiological indices such as TCP and NTCP were calculated from DVH data based on Kehwar et al. using an indigenously developed Matlab program [18]. The software program was cross-validated with an independent Excel calculation. Parameters for bladder and rectum from Emami et al. data were used as the input to calculate NTCP values.

2.1. Statistical Analysis

The values for the dosimetric and radiobiological indices were calculated using paired Student’s test using statistics computation environment STATISTICA 5.0 (StatSoft Inc., USA). Confidence intervals were evaluated using the values. Pearson product moment correlation method was used to correlate CI with TCP estimates. The correlation coefficient was computed. The 2D scatter plot demonstrating the correlation was generated.

2.2. Results

G_O resulted in a higher CI with a mean value of 0.8879 compared to the graphical optimized plan (). EI was less in Gr_O plans (mean = 0.0009). Homogeneity of dose distribution is higher for Gr_O and DP_N plans which is obvious from the mean DHI values 0.6629 and 0.6282, respectively, whereas other methods resulted in a lesser homogeneous dose inside the target (Table 1). ODI and DNR were less for Gr_O and DP_N methods. The dose to 2cc volumes of bladder and rectum were minimized with Gr_O method (B2cc = 4.25 Gy; R2cc = 4.16 Gy). Dose point normalization resulted in a lesser bladder dose (mean 4.61 Gy) which is comparable to Gr_O plans. TCP computed was found to be similar for all the five planning techniques with no statistically significant results. NTCP for bladder (mean: 0.3252%) and rectum was less (mean: 0.3126%) for Gr_O plans whereas G_O plan resulted in high NTCP values (Table 1).

Dose distributions resulted from each method are analysed using the axial CT sections (Figure 1). The high dose volumes (>100% of the prescribed dose) were found to be less for Gr_O and DP_N plans compared to other plans. Gr_O plan had smaller volumes of higher isodose regions (150% and 200%) only around the needles. Figure 2 shows the scatter plot diagram of the Pearson product moment correlation between CI and TCP. The regression coefficient () computed was −0.120 which is not statistically significant (confidence interval: 95%).

125020.fig.001
Figure 1: Comparative axial slice view of the ISBT plans.
125020.fig.002
Figure 2: Graph correlating CI and TCP.
2.3. Discussion

HDR BT often results in heterogeneous dose distribution owing to the steep dose falloff in the vicinity of the radioactive source. But a uniform dose is required to achieve the optimal tumor cell killing avoiding the necrosis. Normalization and optimization algorithms are being used for dose calculation in CT based planning of cervix implants. These methods are said to improve the dose homogeneity within the tumor. Geometrical optimization of dose distributions depends only on the implant volume. A template guided interstitial implant procedure done without proper image guidance could encompass the normal structures. Hence it increases the risk of normal tissue complications. In dose point optimization, reference isodose line is optimized along predefined dose points (interspace distance: 3 mm) on the target’s surface. Bladder and rectal volumes are proximal to the delineated cervix. Hence dose optimized to surface of target volume encompasses the normal structures. graphical optimization helps in dragging the isodose lines that resulted from dose point optimization reducing critical organ doses. Gr_O is a powerful optimization tool which allows the planner to drag the isodose line to a desired level. Thus dwell weights/times are adjusted accordingly to meet the dose constraints. Appropriate and efficient use of the tool is still a learning curve for the planners. Efficiency of optimization and normalization techniques needs to be quantified to choose the optimal one for achieving better clinical outcomes.

The current research report that is the first one to evaluate interstitial brachytherapy plans with dosimetrical and radiobiological indices (TCP and NTCP) was performed for a larger set of DVH data (). Tumor control and normal tissue complication probability estimation using linear quadratic formula devised by Kehwar et al. for interstitial implants were used. Voxel element calculations were used to account for high dose gradients in the HDR implants. A better dose homogeneity can be achieved with the graphically optimized and volume normalized plans compared to others. But the normalization around the implant geometry (geometrical normalization) had resulted in more inhomogeneous dose distribution (mean DHI = 0.4494). Target dose homogeneity was maintained in Gr_O plans. Overdose Volume Index (ODI) and Dose Non uniformity Ratio (DNR) showed no statistically signiicant diference between Gr_O and DP_N methods. The External Volume Index quantifying the normal tissue dose outside target volume is very less for graphically optimized plans with an extreme statistical significance with other methods. Bladder and rectal volumes usually remain very close to the target volume in cervical cancer. Conformity Index was compromised with the manual pulling of isodose lines from the critical organs which typically engross the target volume. The dosimetric analysis of target coverage with dose volume parameters may imply inadequate target coverage with graphical optimization. But it may not be influenced by the tumor control probability since there was no correlation with Conformity Index. TCP estimates were comparable to those of normalization and optimization methods. Thus the use of graphical optimization in increasing tumor control and reducing bladder and rectal toxicities can be established.

CI and TCP cannot be correlated because of the statistically insignificant regression coefficient (Pearson product moment correlation). The values computed for the TCP estimates had no statistical difference. When the dose point (surface) optimized dose distributions are graphically or manually optimized, the dose to the 2cc volume of bladder and rectum can be reduced well compared to other automated dose calculation methods with extremely good statistical significance (95% confidence interval). Fairly good normal tissue complication probability estimates were also obtained for Gr_O plans. The use of graphical optimization in planning the interstitial cervix cases can spare the bladder and rectal doses to a greater extent without compromising the TCP estimates. Grade I and Grade II bladder and rectal complexities were observed in the clinical follow-ups (on treatment and posttreatment).

3. Conclusions

graphical optimization is more beneficial in reducing radiotherapy side effects without compromising disease control. Hence it can be suggested as a method of choice for planning the interstitial cervix implant cases. The potential dosimetric results need to be corroborated with clinical outcomes.

Abbreviations

CT:Computed tomography
HDR:High dose rate
ISBT:Interstitial brachytherapy
Gr_O:Graphical optimization
V_N:Volume normalization
V_O:Volume optimization
G_N:geometrical normalization
G_O:Geometrical optimization
CI:Conformity Index
EI:External Volume Index
DHI:Relative Dose Homogeneity Index
ODI:Overdose Volume Index
DNR:Dose Nonuniformity Ratio
Bl2cc :Dose to 2cc volume of bladder
R2cc:Dose to 2cc volume of rectum
TCP:Tumor control probability
NTCP:Normal tissue complication probability.

Conflict of Interests

The authors declare that there is no conflict of interests regarding the publication of this paper.

References

  1. A. Nandakumar, T. Ramnath, and M. Chaturvedi, “The magnitude of cancer cervix in India,” Indian Journal of Medical Research, vol. 130, no. 3, pp. 219–221, 2009. View at Scopus
  2. G. A. Viani, G. B. Manta, E. J. Stefano, and L. I. de Fendi, “Brachytherapy for cervix cancer: low-dose rate or high-dose rate brachytherapy—a meta-analysis of clinical trials,” Journal of Experimental and Clinical Cancer Research, vol. 28, no. 1, article 47, 2009. View at Publisher · View at Google Scholar · View at Scopus
  3. H. M. Keys, B. N. Bundy, F. B. Stehman et al., “Cisplatin, radiation, and adjuvant hysterectomy compared with radiation and adjuvant hysterectomy for bulky stage IB cervical carcinoma,” The New England Journal of Medicine, vol. 340, no. 15, pp. 1154–1161, 1999. View at Publisher · View at Google Scholar · View at Scopus
  4. B. P. C. Goksedef, C. Kunos, J. L. Belinson, and P. G. Rose, “Concurrent cisplatin-based chemoradiation International Federation of Gynecology and Obstetrics stage IB2 cervical carcinoma,” American Journal of Obstetrics and Gynecology, vol. 200, no. 2, pp. 175.e1–175.e5, 2009. View at Publisher · View at Google Scholar · View at Scopus
  5. F. B. Stehman, S. Ali, H. M. Keys et al., “Radiation therapy with or without weekly cisplatin for bulky stage 1B cervical carcinoma: follow-up of a Gynecologic Oncology Group trial,” American Journal of Obstetrics and Gynecology, vol. 197, no. 5, pp. 503.e1–503.e6, 2007. View at Publisher · View at Google Scholar · View at Scopus
  6. P. J. Eifel, K. Winter, M. Morris et al., “Pelvic irradiation with concurrent chemotherapy versus pelvic and para-aortic irradiation for high-risk cervical cancer: an update of Radiation Therapy Oncology Group Trial (RTOG) 90-01,” Journal of Clinical Oncology, vol. 22, no. 5, pp. 872–880, 2004. View at Publisher · View at Google Scholar · View at Scopus
  7. R. Pearcey, M. Brundage, P. Drouin et al., “Phase III trial comparing radical radiotherapy with and without cisplatin chemotherapy in patients with advanced squamous cell cancer of the cervix,” Journal of Clinical Oncology, vol. 20, no. 4, pp. 966–972, 2002. View at Publisher · View at Google Scholar · View at Scopus
  8. C. W. Whitney, W. Sause, B. N. Bundy et al., “Randomized comparison of fluorouracil plus cisplatin versus hydroxyurea as an adjunct to radiation therapy in stage IIB-IVA carcinoma of the cervix with negative para-aortic lymph nodes: a Gynecologic Oncology Group and Southwest Oncology Group Study,” Journal of Clinical Oncology, vol. 17, no. 5, pp. 1339–1348, 1999. View at Scopus
  9. P. K. Nandwani, R. K. Vyas, J. P. Neema, U. K. Suryanarayan, D. C. Bhavsar, and K. Jani, “Retrospective analysis of role of interstitial brachytherapy using template (MUPIT) in locally advanced gynecological malignancies,” Journal of Cancer Research and Therapeutics, vol. 3, no. 2, pp. 111–115, 2007. View at Scopus
  10. C. Haie-Meder, A. Gerbaulet, and R. Potter, The GEC ESTRO Handbook of Brachytherapy, chapter 17, Interstitial Brachytherapy in Gynecological Cancer, 2002.
  11. S. S. Supe, K. M. Ganesh, H. Vaithianathan, and B. N. Sankar, “Radiobiological quality of high dose rate interstitial brachytherapy treatments of carcinoma of the cervix,” Reports of Practical Oncology and Radiotherapy, vol. 11, no. 1, pp. 13–22, 2006. View at Scopus
  12. B. Shwetha, M. Ravikumar, A. Katke et al., “Dosimetric comparison of various optimization techniques for high dose rate brachytherapy of interstitial cervix implants,” Journal of Applied Clinical Medical Physics, vol. 11, no. 3, pp. 225–230, 2010. View at Scopus
  13. S. Jamema, S. Saju, U. Shetty, S. Pallad, D. Deshpande, and S. Shrivastava, “Dosimetric comparison of inverse optimization with geometric optimization in combination with graphical optimization for HDR prostate implants,” Journal of Medical Physics, vol. 31, no. 2, pp. 89–94, 2006. View at Scopus
  14. H. Meertens, J. Borger, M. Steggerda, and A. Blom, “Evaluation and optimization of interstitial brachytherapy dose distribution,” in Brachytherapy from Radium to Optimization, R. F. Mould, J. J. Battermann, A. A. Martinez, and B. L. Speiser, Eds., pp. 300–306, Nucletron International, Veenendaal, The Netherlands, 1994.
  15. “HDR micro Selectron V2 Treatment planning BPS version 14. 2/14.3—User manual,” Nucletron, The Netherlands.
  16. A. Wu, K. Ulin, and E. S. Sternick, “A dose homogeneity index for evaluating 192Ir interstitial breast implants,” Medical Physics, vol. 15, no. 1, pp. 104–107, 1988. View at Publisher · View at Google Scholar · View at Scopus
  17. V. Chaswal, S. Yoo, B. R. Thomadsen, and D. L. Henderson, “Multi-species prostate implant treatment plans incorporating 192Ir and 125I using a Greedy Heuristic based 3D optimization algorithm,” Medical Physics, vol. 34, no. 2, pp. 436–444, 2007. View at Publisher · View at Google Scholar · View at Scopus
  18. T. S. Kehwar, S. F. Akber, and K. Passi, “Qualitative dosimetric and radiobiological evaluation of high—dose—rate interstitial brachytherapy implants,” International Journal of Medical Sciences, vol. 5, no. 1, pp. 41–49, 2008. View at Scopus