Clinical Study
Clinical Evaluation of a Mobile Heart Rhythm Telemonitoring System
Table 2
Correlations and intraclass correlation coefficients (ICC) between different parameters measured with standard Holter ECG and TEMEO monitoring system.
| Parameter | ECG Holter | TEMEO system | Correlation | ICC | P value |
| Average HR | 75 ± 14 | 74 ± 16 | 0.85 | 0.91 | <0.001 | Maximal HR | 130 ± 33 | 121 ± 19 | 0.42 | 0.48 | 0.02 | Minimal HR | 54 ± 14 | 48 ± 13 | 0.77 | 0.84 | <0.001 | % time in tachycardia | 10.2 ± 16.3% | 9 ± 15.1% | 0.59 | 0.72 | <0.001 | VEB single | 456 ± 618 | 363 ± 722 | 0.43 | 0.58 | 0.02 | VEB couplets | 109 ± 258 | 76 ± 195 | 0.75 | 0.47 | 0.01 | SVEB single | 815 ± 1002 | 315 ± 678 | 0.68 | 0.59 | 0.001 | SVEB couplets | 105 ± 218 | 33 ± 106 | 0.19 | 0.24 | 0.05 |
|
|
HR: heart rate; VEB: ventricular ectopic beats; SVEB: supraventricular ectopic beats.
|