Clinical Study

Clinical Evaluation of a Mobile Heart Rhythm Telemonitoring System

Table 2

Correlations and intraclass correlation coefficients (ICC) between different parameters measured with standard Holter ECG and TEMEO monitoring system.

ParameterECG HolterTEMEO systemCorrelationICCP value

Average HR75 ± 1474 ± 160.850.91<0.001
Maximal HR130 ± 33121 ± 190.420.480.02
Minimal HR54 ± 1448 ± 130.770.84<0.001
% time in tachycardia10.2 ± 16.3%9 ± 15.1%0.590.72<0.001
VEB single456 ± 618363 ± 7220.430.580.02
VEB couplets109 ± 25876 ± 1950.750.470.01
SVEB single815 ± 1002315 ± 6780.680.590.001
SVEB couplets105 ± 21833 ± 1060.190.240.05

HR: heart rate; VEB: ventricular ectopic beats; SVEB: supraventricular ectopic beats.