Clinical Study

Clinical Evaluation of a Mobile Heart Rhythm Telemonitoring System

Table 3

Correlations and intraclass correlation coefficients (ICC) between maximal and minimal HR measured with standard Holter ECG holter, TEMEO monitoring system, and manually.

ParameterECG HolterManual TEMEO systemCorrelationICCP value

Maximal HR133 ± 33113 ± 230.30.440.014
113 ± 23121 ± 190.680.8<0.001

Minimal HR55 ± 1456 ± 110.90.93<0.001
56 ± 1148 ± 130.720.83<0.001

HR: heart rate.