Spectrophotometric Estimation of Flutamide in Pure and in Pharmaceutical Preparations
Table 1
Optical characteristics and precision.
Parameter
Method A
Method B
Method C
nm
410
440
500
Linear range (μg/mL)
0.5–12
0.5–10
0.3–10
Molarabsorptivity, (L/mol/cm)
2.23 104
2.04 104
1.35 104
Sandell sensitivity (μg cm−2)
0.0281
0.0136
0.0205
Intercept ()
0.0124
0.0179
0.0116
Slope ()
0.0734
0.0669
0.0439
Correlation coefficient ()
0.9988
0.9976
0.9975
0.0182
0.0248
0.0167
0.0012
0.0027
0.0018
LOQ (μg/mL)
0.1197
0.1047
0.2587
LOD (μg/mL)
0.0395
0.0345
0.0854
* where is the concentration of flutamide in μg/mL and is the absorbance at the respective . is the standard deviation of intercept, is the standard deviation of slope.