Research Article

Development and Validation of RP-HPLC Method for Azilsartan Medoxomil Potassium Quantitation in Human Plasma by Solid Phase Extraction Procedure

Table 1

Results of 5 calibration curves for AMP determination.

Added conc.123456789

Back calculated conc. ( g/mL)1.011.983.104.384.976.086.828.059.08
1.041.953.084.005.005.587.227.908.85
1.071.933.073.874.946.117.137.818.85
1.032.012.994.195.376.467.428.369.67
1.082.063.083.554.706.016.767.528.92

Mean1.051.983.064.005.006.057.077.939.07
Mean accuracy%104.5399.23102.1599.9499.94100.81100.9899.10100.81
RSD%2.552.591.407.844.775.203.913.943.83