Research Article

Development and Validation of a Simple and Sensitive Spectrometric Method for Estimation of Cisplatin Hydrochloride in Tablet Dosage Forms: Application to Dissolution Studies

Table 3

Method accuracy results for cisplatin hydrochloride tablets.

Sample ( g/mL) cisplatin hydrochlorideReference standard concentration added ( g/mL)Total concentration of solution ( g/mL)Concentration of drug found ( g/mL)Recovery (%)R.S.D. (%) ( )Mean recovery (%)

0.80.41.21.1999.160.16 99.98
0.81.61.61100.621.2
1.22.02.0100.180.92