Research Article

Determination of Magnesium Valproate and Its Process Related Impurities by Ultraperformance Liquid Chromatography

Table 2

Summary of validation study.

Validation experimentPA2-EPA2-1-MEPAVAMgVA

SpecificityNo interferenceNo interferenceNo interferenceNo interferenceNo interference
LOQ ( g/mL)2.02.02.02.02.0
LOD ( g/mL)1.01.01.01.01.0
Linearity
(I) Correlation coefficient0.9990.9980.9990.9970.998
(II) Regression equation
Method precision ( )%RSD = 0.2%RSD = 0.2%RSD = 0.2%RSD = 1.2%RSD = 0.1
Int. precision ( )%RSD = 1.2%RSD = 0.8%RSD = 0.8%RSD = 1.8%RSD = 0.7
Accuracy (% recovery)99–10198–10099-10098–10299-100
RobustnessNo significant changeNo significant changeNo significant changeNo significant changeNo significant change