Research Article
Validation of a Stability-Indicating RP-HPLC Method for Determination of L-Carnitine in Tablets
Table 2
Precision, accuracy, and recovery data for the proposed method.
| L-Carnitine concentration (µg/mL) | Precision (RSD, %) |
Accuracy () mean (RSD, %) | Recovery () | Intraday () | Interday () | Target concentration (mg/mL) | Calculated concentration mg/mL (mean) | Percentage | ±SD |
| 84.74 | 3.34 | 2.18 | 97.27 (2.38) | 2.20 | 2.22 | 100.91 | 0.04 | 169.48 | 0.41 | 1.36 | 100.72 (0.75) | 2.40 | 2.42 | 100.83 | 0.01 | 1355.80 | 0.83 | 0.99 | 98.52 (0.83) | 2.60 | 2.64 | 101.54 | 0.02 | 3389.50 | 1.84 | 1.60 | 99.51 (1.05) | | | | |
|
|