Research Article

Stability Indicating RP-HPLC Method for the Simultaneous Determination of Atorvastatin Calcium, Metformin Hydrochloride, and Glimepiride in Bulk and Combined Tablet Dosage Form

Table 1

System suitability parameters of ATR, MET, and GLM.

ParameterMETATRGLM Specifications [22]
Mean ± SD%RSDMean ± SD%RSDMean ± SD%RSD

Rt2.57 ± 0.041.757.06 ± 0.131.909.39 ± 0.111.23RSD ≤ 2
TF1.25 ± 0.010.921.15 ± 0.011.001.11 ± 0.010.89TF ≤ 2
Rs9.38 ± 0.181.925.27 ± 0.061.18
TP 3556 ± 641.824872 ± 931.926280 ± 300.489 > 2000

Rt; retention time, TF: tailing factor, Rs: resolution, and TP: theoretical plates; .