Abstract
The newly developed QA-810V is an optional unit for the determination of five-part white blood cell differentials. It can be used together with the same manufacturer's haematology analyser which has been used in relatively small-sized laboratories. The present study evaluates the basic performance of the QA-810V and the MEK-8118 haematology analyser using routinely obtained blood specimens treated with ethylenedioaminetetraacetic acid-2K. In this evaluation, reproducibility was good and little carryover was found. Accurate measurements were possible for up to 24h of storage. Storage at 4°C yielded more stable measurements of complete blood counts and five-part differentials than storage at room temperature. A good correlation between findings with the MEK-8118 haematology analyser and those with the SE-9000 haematology analyser was found for complete blood counts. The leukocyte differential obtained with the QA-810V correlated well with eye counts, with