Research Article
RP-HPLC Method Development and Its Validation for Quantitative Determination of Rimonabant in Human Plasma
Table 1
Imprecision and inaccuracy results of validation.
| Spiked conc. (μg/L) | Conc. found mean ± SD (μg/L) | CV (%) | Accuracy (%) |
| Intraday imprecision and inaccuracy () | 150 | | 4.0 | 101.2 | 450 | | 4.4 | 98.3 | 800 | | 3.2 | 97.6 | Interbatch imprecision and inaccuracy () | 150 | | 3.4 | 102.6 | 450 | | 3.9 | 95.7 | 800 | | 3.7 | 101.3 | Interday imprecision and inaccuracy () | 150.0 | | 5.1 | 104.3 | 450.0 | | 4.1 | 96.7 | 800.0 | | 2.7 | 94.5 | Analyst to analyst imprecision and inaccuracy () | 150.0 | | 6.4 | 106.7 | 450.0 | | 2.1 | 101.8 | 800.0 | | 3.2 | 97.5 | Global imprecision and inaccuracy () | 150.0 | | 4.3 | 101.9 | 450.0 | | 5.1 | 97.0 | 800.0 | | 4.6 | 99.5 |
|
|