Research Article

Quantification of Neurotransmitters in Mouse Brain Tissue by Using Liquid Chromatography Coupled Electrospray Tandem Mass Spectrometry

Table 2

Determination of neurotransmitters by LC-MS/MS: validation results on precision and accuracy.

ChemicalIntraday precisionaInterday precisionbAccuracy (%)RE (%)
LowMidHighLowMidHighLowMidHighLowMidHigh

BH41.181.090.424.281.780.59102.9099.7297.482.900.282.53
Dopamine1.151.260.733.981.900.86103.0799.1297.803.070.882.21
Serotonin8.308.128.208.589.449.2395.57108.4696.154.438.463.85
Norepinephrine8.216.954.5010.177.284.93104.90103.0698.754.903.061.26
Epinephrine6.224.512.456.887.502.5793.0093.9097.917.006.102.09
Glutamate3.134.183.163.634.263.41101.0097.1095.811.002.904.19
GABA2.763.773.473.704.154.06103.50101.2793.943.501.276.06

For BH4, Dopamine, Serotonin, norepinephrine, epinephrine, the low, medium, and high control solutions were 30 pg/mg, 500 pg/mg, and 8000 pg/mg of DW, respectively. For glutamate and GABA, those were 6000 pg/mg, 30,000 pg/mg, and 160,000 pg/mg of DW.
For both precision tests, the values were in coefficient of variation (CV).
RE = Relative Error.
Mean of five replicates () observations at each concentration.
bMean of 25 replicates () observations over five different analytical runs.