Research Article

Development of Sensitive and Specific Analysis of Vildagliptin in Pharmaceutical Formulation by Gas Chromatography-Mass Spectrometry

Table 3

Intra- and interday precision and accuracy results for VIL ().

Nominal amount (ng mL−1)Interday Intraday
  ± SEPrecision (%)Accuracy (%)  ± SEPrecision (%)Accuracy (%)

1515.31 ± 0.193.622.0615.2 ± 0.223.041.33
150150.7 ± 1.682.730.46151.16 ± 1.492.410.77
250249.3 ± 1.871.84−0.26251.01 ± 1.201.170.40

  ± SE: mean ± standard error.
Percent relative standard deviation.
Bias: [(found − added)/added] × 100.