Research Article

A Study on the Reliability of an On-Site Oral Fluid Drug Test in a Recreational Context

Table 2

Linearity of the HS-SPME-GC-MS procedure for compounds under investigation.

CompoundsSlopeInterceptLOD (ng/pad)LOQ (ng/pad)Analytical recovery (%)

A0.640.070.9950.371.11100.8
MA2.080.200.9900.682.0499.2
MDA1.93−0.160.9880.601.8088.8
Ketamine6.47−1.160.9900.732.1984.4
MDMA5.570.860.9960.341.02100.8
MDE5.730.140.9980.260.78103.2
MBDB8.340.770.9990.110.3398.8
Methadone9.74−1.460.9900.712.1385.5
Cocaine6.180.050.9960.351.05107.6
Cocaethylene2.01−0.420.9920.511.5390.4
THC10.520.030.9900.060.18102.0
Codeine4.730.630.9960.060.1896.0
Morphine7.322.050.9950.431.2996.4
6-MAM9.523.740.9930.391.17105.6

A, amphetamine; MA, methamphetamine; MDA, methylenedioxyamphetamine; MDMA, methylenedioxymethamphetamine; MDE, methylenedioxyethamphetamine; MBDB, N-methyl-1-(1,3-benzodioxol-5-yl)-2-butanamine; THC, 9-Tetrahydrocannabinol; 6-MAM, 6-monoacetylmorphine.