Research Article

Quality Control of 11 Cannabinoids by Ultraperformance Liquid Chromatography Coupled with Mass Spectrometry (UPLC-MS/MS)

Table 1

Statistical analysis of a six-point calibration curve from 1 ng/mL to 200 ng/mL, with five replicates for calibration standards and quality control (QC) standards for the 11 cannabinoid mixture. Tandem mass spectrometry (MS/MS) transitions calibration range results for potency. The table shows the quantitative abilities of this method for each analyte.

NamesRetention time (min)MS/MS ion transitionsConcentration ranges (ng/mL)Linearity (R2)LODLOQ

Δ9-THCA6.4357.2101 ➔ 313.21450–2000.9901025
CBDA4.3357.2066 ➔ 245.15380–2000.995510
CBGA4.5359.2192 ➔ 341.21260–2000.992510
CBG4.5317.2470 ➔ 193.12230–2000.9931050
CBD4.6315.2336 ➔ 259.16680–2000.9961025
THCV4.5287.2031 ➔ 165.09290–2000.9981025
CBN5.3311.2011 ➔ 223.11300–2000.997510
Δ8-THC5.91315.2336 ➔ 193.12420–2000.993510
Δ9-THC5.8315.2324 ➔ 259.17050–2000.9921025
CBC6.4315.2336 ➔ 193.12420–2000.9912550
CBDV3.4287.19771 ➔ 65.08920–2000.992510

LOD, limit of detection; LOQ, limit of quantification. Values of linearity refer to the linear range.