Analytical figures of merit for the determination of Cd and Pb in swimming pool waters by GF AAS after preconcentration on polyurethane foam.

ParameterPb(II)Cd(II)

Concentration range1.8–40 μg·L−10.06–3 μg·L−1
 LOD0.5 μg·L−10.02 μg·L−1
 LOQ1.8 μg·L−10.06 μg·L−1
Calibration equationA = 0.0069 (Pb (μg L−1)) + 0.0047A = 0.16 (Cd (μg L−1)) + 0.0018
r20.99760.9996
 Intermediary precision7.3%6.1%