Portable XRF Technology to Quantify Pb in Bone In Vivo

Table 2

Phantom Pb concentrations calculated using the background subtraction method.

Standard phantom ppmLucite thickness
0 mm1 mm2 mm3 mm

0−0.38 ± 0.8−0.1 ± 1.3−2.22 ± 2.13.31 ± 3.31
54.83 ± 0.863.57 ± 1.349.61 ± 2.197.37 ± 3.37
1010.23 ± 0.910.5 ± 1.3911.23 ± 2.1917.83 ± 3.42