Research Article

High Throughput Quantitative Bioanalytical LC/MS/MS Determination of Gemifloxacin in Human Urine

Table 1

Data of back-calculated GFX concentration of the calibration standards in human urine.

Nominal concentration (ng mL−1)MeanaStandard deviation (SD)Precision (%)Accuracy (%)

54.970.122.4399.43
1514.780.241.5998.54
2525.530.622.43102.11
5052.351.843.52104.69
150156.753.182.03104.50
200205.594.342.11102.79
400392.392.730.6998.09
500502.412.330.46100.48

Average of six determinations.