Stability-Indicating Photochemical Method for the Assay of Riboflavin: Lumichrome Method

Table 1

Validation data for multicomponent spectrometric assay of RF, LC, and LFa.

CompoundRFLCLF

nm (pH 2.0)445356445
Molar absorptivity () M−1 cm−1125001080010400
Linearity
 Concentration range (M)0.5–5.0 × 10−50.5–5.0 × 10−50.5–5.0 × 10−5
 Slope ×10−41.2501.0801.040
 Intercept0.00260.00440.0035
 SD of slope±0.003±0.003±0.005
 SE of slope0.9681.4151.025
Recovery range (%)98.5–101.298.2–100.698.0–101.8
Accuracy (%) ± SD100.2 ± 1.2199.8 ± 1.5899.7 ± 1.50
RSD (%)1.201.501.59
LOD (M)1.195 × 10−61.251 × 10−62.160 × 10−6
LOQ (M)4.183 × 10−64.405 × 10−64.420 × 10−6

Values are mean of five determinations.