Stability-Indicating Photochemical Method for the Assay of Riboflavin: Lumichrome Method

Table 4

Assay of RF in photodegraded samples and aged vitamin preparations (25 ± 1°C).

MethodSample stored in transparent containers (pH 7.0)aVitamin preparations stored in original package (pH 3-4)b
Added (mg/100 mL)Found (mg/100 mL)% recovery% RSDLabeled (mg/100 mL)Found (mg/100 mL)% recovery% RSD

USP fluorimetric methodc
 Sample 133.221.4664.631.3933.230.3191.291.33
 Sample 224.015.1663.161.9724.022.4093.331.75
 Sample 350.035.4270.841.5550.044.7589.501.45
LC methodc
 Sample 133.220.8862.891.9733.229.6389.241.98
 Sample 224.014.8061.661.9724.022.3092.911.81
 Sample 350.035.3170.621.9650.043.9987.981.88

Light for 4 h.
bRoom light for 6 months.
cThe tabulated value at the 95% confidence level for five degrees of freedom is 2.571 and the calculated value is 1.950. Therefore, < and there is no significant difference between the two methods at this confidence level.