Research Article

Fractional-Order Control of a Nonlinear Time-Delay System: Case Study in Oxygen Regulation in the Heart-Lung Machine

Table 2

Indices of the system time response due to using 𝑃 𝐼 𝐷 and 𝑃 𝐼 𝜆 𝐷 𝜇 controllers.

(Z-N)IAEISEITAE
𝑃 𝐼 𝐷 𝑃 𝐼 𝐷 𝑃 𝐼 𝜆 𝐷 𝜇 𝑃 𝐼 𝐷 𝑃 𝐼 𝜆 𝐷 𝜇 𝑃 𝐼 𝐷 𝑃 𝐼 𝜆 𝐷 𝜇

Rise time245 sec98 sec45 sec35 sec26 sec115 sec31 sec
Settling time1783 sec300 sec100 sec612 sec565 sec300 sec115 sec
Overshot %50%16.8%18%26.18%26.5%10.12%28%
Reached criteria value9816.34626.4769840327900868350465820