Experimental Investigations on Microcracks in Vibrational and Conventional Drilling of Cortical Bone

1Key Laboratory for Biomechanics and Mechanobiology of Ministry of Education, International Joint Research Center of Aerospace Biotechnology and Medical Engineering, Ministry of Science and Technology of China, School of Biological Science and Medical Engineering, Beihang University, Beijing 100191, China
2Research Institute of Beihang University in Shenzhen, Shenzhen 518057, China

Received 14 June 2013; Revised 17 July 2013; Accepted 18 July 2013

Academic Editor: Xiaoming Li

Copyright © 2013 Yu Wang et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Bone drilling is widely used in orthopedic surgery. Microcracks will be generated in bone drilling, which may cause fatigue damages and stress fractures. Fresh bovine cortical bones were drilled via vibrational and conventional ways. Drilling operations were performed by a dynamic material testing machine, which can provide the vibration while maintaining uniform feed motion. The drill site and bone debris were observed through scanning electron microscope (SEM). The experimental results showed that fewer and shorter micro-cracks were formed in vibrational drilling than those formed in conventional way. And the surface morphology of bone debris from two different drilling ways was also quite different. It is expected that vibrational drilling in orthopedic surgery operation could decrease the microdamage to the bone, which could lower the incidence of stress fracture and contribute to the postoperative recovery.

1. Introduction

Recently, minimally invasive surgery has become one of the most important trend of orthopedic surgery, and it could decrease surgical trauma and shorter the recovery time for patients. As one of the most basic and common surgical operations, bone drilling has been widely used in the orthopedic surgery [1]. Drilling operation itself could also cause damages to the bone. Recent research in this area focused on the thermal damage in bone drilling, which could lead to bone necrosis or even osteomyelitis [2, 3]. In vivo animal experiments showed that if the temperature of bone rises to 50°C for more than one minute, it will cause the cortical bone necrosis [46]. Lundskog and Krause showed that when the temperature of bone rises above 50°C, thermal effects will lead to the degeneration of enzymes and membrane proteins in the bone, which causes bone necrosis [7, 8].

On the other hand, high-speed drilling also makes damages to bone microstructure. Experimental studies have indicated that many microcracks were generated in bone drilling. Meanwhile, excessive microcracks will be produced when the bone strain exceeds the threshold. These microcracks may cause fatigue damages in bone, because the increase of microcracks can decrease the bone stiffness and the elastic modulus [912]. Only a few of these microcracks can be repaired by bone remodeling process [1315]. Therefore, rapid accumulation of microcracks will increase the bone fragility and lead to stress fractures [16]. When microcracks accumulate at normal rate and the bone’s repair mechanism is deficient, it will lead to the collapse of the bone and even fragility fractures [17, 18].

Vibrational drilling is a new drilling method, which adds axial vibration to the drill bit. Additional vibrational pulse can effectively improve the drilling effect and reduce the cutting heat in the drilling process [19]. Alam et al. reported that vibrational drilling could significantly reduce drilling force and torque in cortical bone drilling [20]. Both cutting heat and drilling force/torque represent drilling energies, which have relationship with bone microcracks. Therefore, in this work, the hypothesis is that vibrational drilling can decrease the accumulation in microcracks during bone drilling. We focused on the efforts to study the damages of vibrational and conventional drilling methods on bone microstructure. The differences in microcracks for different drilling methods were observed by the scanning electron microscope (SEM). This comparison research of microcracks in different drilling methods may suggest a new way to reduce the microcracks in bone drilling by changing drilling methods.

2. Materials and Methods

2.1. Specimen Preparation

According to current studies, the properties of bovine bones best resemble the properties of human bone [21]. Therefore cortical bone specimens, which were cut from bovine femur, were used in this research. Considering that the thickness of the cortical bone is not completely uniform, the bovine femur was cut into 1 cm long sections by a precision cutting machine (SYJ-200, China; Figure 1). Taking into account the average thickness of the cortical bone and to ensure the same drilling depth in the experiment, the drilling depth is fixed to 8 mm. The marrow was removed, and cortical bone thickness over 8 mm was marked for drilling as shown in Figure 2. Then the specimens were kept frozen until used. One hour before each experiment, the specimens were thawed in a thermostat-controlled bath adjusted at room temperature to guarantee each of the initial specimens was at room temperature before drilling.

845205.fig.001
Figure 1: The precision cutting machine.
845205.fig.002
Figure 2: Specimen for drilling.
2.2. Experimental Setup

As shown in Figure 3, in vitro drilling operations were performed by a dynamic material testing machine (Instron E10000, USA), which can provide the vibration while maintaining uniform feed motion (  mm/min). A vise was located on the table of the machine centre to fix the specimens. A hand drill was connected with a flexible drive rod, which can transfer rotation to another drill bit. The drill bit was fixed to the testing head of the machine with stainless steel drill on it. In this research, the drill speed is fixed at 8000 rad/min, and the diameter of the drill bit is 4 mm. It is well known that cutting heat is generated from friction and internal structural damages. And the temperature was considered as an important index to reflect the drilling energy. Therefore, temperature changes of drill holes were directly measured by Pt100 platinum resistors once finishing drilling.

845205.fig.003
Figure 3: Experimental arrangement for vibrational and conventional drilling.

Before drilling, the prepared sample which has been adjusted to room temperature was fixed to the vise. Then the drill bit was aligned with the predrilled site by manual controlling. Afterward, the drilling parameters such as drilling depth and feed rate were set in the software which comes with the dynamic material testing machine. In the end, we turned on the hand drill and ran the software. When the drilling finished, we turned off the hand drill and inserted the Pt100 platinum resistors into the drill hole to measure the temperature of the drill hole.

Conventional and vibrational drillings with different parameter sets (frequency: 5~20 Hz, amplitude: 100~500 μm) were performed. The drill site of cortical bone and bone debris from conventional drilling group and the vibrational drilling group with the lowest temperature ( μm,  Hz) were observed through SEM (CamScan 3400, German).

3. Results

In this study, a total of 60 experiments were conducted in two drilling methods, which contained 1 conventional drilling group and 5 vibrational drilling groups. 10 drillings were carried out for each group. The mean temperatures of drill holes for every group were summarized in Table 1.

tab1
Table 1: The mean temperature (°C) of drill hole and the value for every group.

We can see from Table 1 that there is a downward trend in the mean temperature of drill hole with the vibration increase. Moreover, compared to conventional drilling, vibrational drilling can significantly reduce the cutting heat in drilling of cortical bone ( ).

In order to show the microcracks in bone via different drilling operations, the drilled specimens in conventional drilling group and the fifth vibrational drilling group ( μm,  Hz), which hold the highest and the lowest drill hole temperature, respectively, were observed by SEM.

The specimens were cut along the axis of the drilling hole, and then the drilling site and bone debris, which were collected during drilling, were sprayed with gold and observed with SEM. The microcracks in the SEM pictures were sketched manually using medical image analysis software Digimizer. Table 2 shows the percentages of microcracks area in the SEM pictures from each group with three different magnifications. The SEM results show that the microcracks in vibrational drilling are significantly less than those in conventional drilling. Figure 4 shows two typical scanning pictures from the two groups. The left one is from vibrational drilling group and the right one is from the conventional drilling group.

tab2
Table 2: The percentages of microcracks area in the SEM picture from each group.
fig4
Figure 4: Scanning pictures with magnification 100, left: vibrational drilling; right: conventional drilling.

For a clearer observation of the microcracks in two drilling methods, the magnification was raised to 300 in Figure 5. Similarly, the microcracks were measured manually by Digimizer. In general, it is observed that the length of most microcracks from vibrational drillings is shorter than that from conventional drillings.

fig5
Figure 5: Scanning pictures with magnification 300, left: vibrational drilling; right: conventional drilling.

Meanwhile, the bone debris generated in drilling process was also observed by SEM. It is found that the surface morphology of the bone debris differed in the two groups. Figure 6 is the SEM pictures of bone debris from two groups with magnification of 2000. In conventional drilling group, the gullies on the bone debris surface are very obvious and clear. However, the bone debris surface in vibrational drilling group is relatively flat.

fig6
Figure 6: SEM pictures of bone debris with magnification of 2000, left: vibrational drilling; right: conventional drilling.

4. Discussion

Extensive practice in machining areas has proved that vibrational drilling compared to conventional drilling has obvious advantages [2226]. Studies have shown that the pulse of high-frequency vibration has a significant effect in reducing bone cutting heat [27, 28]. Therefore, the hypothesis is that vibrational drilling may also bring less damage to the bone microstructures and reduce the microcracks generated in bone drillings.

According to Table 1, vibrational drilling can significantly reduce the cutting heat in drilling of cortical bone. We speculate the reason is that vibrational drilling can reduce the contact time of drill bit and drill hole and promote air flow in drill hole, which can carry away part of the cutting heat. On the other hand, the result of SEM showed fewer that micro cracks were formed in vibrational drilling than those formed in conventional way. Therefore, vibrational drilling may also decrease the cutting heat generated by internal structural damage of bone.

Figure 4 shows the scanning pictures with magnification 100 via two drilling operations. From these results it can be observed that drilling method is a critical factor in the generation of microcracks in bone drilling. Compared to the conventional drilling, vibrational drilling has significantly reduced the accumulation in microcracks. As mentioned before, increases in micro-cracking may cause the fatigue damage even to stress fracture. Furthermore, on the terms of the mean length of microcracks, O’Brien et al. reported that microcracks of less than 100 μm in length stopped growing when they encountered a cement line [9]. However, microcracks in the range 150–300 μm continued to grow after encountering a cement line surrounding an osteon. Only microcracks greater than 300 μm were obverted to cause failure [7]. As shown in Figure 5, the mean length of microcracks for vibrational drilling is 100 μm, approximately. But most of the microcracks generated by conventional drilling are about 300 μm in length, which are subject to cause fracture. Therefore, we speculate that vibrational drilling could decrease the incidence of stress fracture.

Figure 6 shows the difference in surface morphology of bone debris. Because the generation of bone debris is related to the energy of drilling, the flat surface in vibrational drilling group may suggest that less energy was transformed to the bone. And we infer that there may be a certain relationship between the surface morphology of the bone debris and the generation of microcracks.

5. Conclusion

According to the experimental results of the microcracks accumulation via vibrational and conventional drilling, it is concluded that vibrational drilling can reduce the generation of microcracks which may contribute to the postoperative recovery.

Conflict of Interests

The authors declare that there is no conflict of interests regarding the publication of this article.

Acknowledgments

This work is financially supported by the Science and Technology Research Foundation of Shenzhen (JC201105190862A), the National Natural Science Foundation (NSFC) Grant of China (61201313), and the 111 Project of China (B13003).

References

  1. K. N. Bachus, M. T. Rondina, and D. T. Hutchinson, “The effects of drilling force on cortical temperatures and their duration: an in vitro study,” Medical Engineering and Physics, vol. 22, no. 10, pp. 685–691, 2000. View at Publisher · View at Google Scholar · View at Scopus
  2. E. T. Berning and R. M. Fowler, “Thermal damage and tracker-pin track infection in computer-navigated total knee arthroplasty,” The Journal of arthroplasty, vol. 26, no. 6, pp. 977–e21, 2011. View at Scopus
  3. X. M. Li, C. A. Van Blitterswijk, Q. L. Feng, F. Cui, and F. Watari, “The effect of calcium phosphate microstructure on bone-related cells in vitro,” Biomaterials, vol. 29, no. 23, pp. 3306–3316, 2008. View at Publisher · View at Google Scholar · View at Scopus
  4. A. R. Eriksson and T. Albrektsson, “Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit,” The Journal of Prosthetic Dentistry, vol. 50, no. 1, pp. 101–107, 1983. View at Scopus
  5. W. Bonfield and C. H. Li, “The temperature dependence of the deformation of bone,” Journal of Biomechanics, vol. 1, no. 4, pp. 323–329, 1968. View at Scopus
  6. X. M. Li, Q. L. Feng, X. H. Liu, W. Dong, and F. Cui, “Collagen-based implants reinforced by chitin fibres in a goat shank bone defect model,” Biomaterials, vol. 27, no. 9, pp. 1917–1923, 2006. View at Publisher · View at Google Scholar · View at Scopus
  7. J. Lundskog, “Heat and bone tissue. An experimental investigation of the thermal properties of bone and threshold levels for thermal injury,” Scandinavian Journal of Plastic and Reconstructive Surgery, vol. 9, pp. 1–80, 1972. View at Scopus
  8. W. R. Krause, “Orthogonal bone cutting: saw design and operating characteristics,” Journal of Biomechanical Engineering, vol. 109, no. 3, pp. 263–271, 1987. View at Scopus
  9. F. J. O'Brien, D. Taylor, and T. C. Lee, “The effect of bone microstructure on the initiation and growth of microcracks,” Journal of Orthopaedic Research, vol. 23, no. 2, pp. 475–480, 2005. View at Publisher · View at Google Scholar · View at Scopus
  10. D. Taylor and J.-H. Kuiper, “The prediction of stress fractures using a “stressed volume” concept,” Journal of Orthopaedic Research, vol. 19, no. 5, pp. 919–926, 2001. View at Publisher · View at Google Scholar · View at Scopus
  11. X. M. Li, Y. Yang, Y. B. Fan, et al., “Biocomposites reinforced by fibers or tubes, as scaffolds for tissue engineering or regenerative medicine,” Journal of Biomedical Materials Research A, 2013. View at Publisher · View at Google Scholar
  12. X. M. Li, Y. Huang, L. S. Zheng, et al., “Effect of substrate stiffness on the functions of rat bone marrow and adipose tissue derived mesenchymal stem cells in vitro,” Journal of Biomedical Materials Research A, 2013. View at Publisher · View at Google Scholar
  13. D. B. Carter, W. E. Caler, D. M. Spengler, and V. H. Frankel, “Fatigue behavior of adult cortical bone: the influence of mean strain and strain range,” Acta Orthopaedica Scandinavica, vol. 52, no. 5, pp. 481–490, 1981. View at Scopus
  14. X. M. Li, L. Wang, Y. B. Fan, et al., “Nanostructured scaffolds for bone tissue engineering,” Journal of Biomedical Materials Research A, vol. 101, no. 8, pp. 2424–2435, 2013. View at Publisher · View at Google Scholar
  15. X. M. Li, H. Gao, M. Uo et al., “Effect of carbon nanotubes on cellular functions in vitro,” Journal of Biomedical Materials Research A, vol. 91, no. 1, pp. 132–139, 2009. View at Publisher · View at Google Scholar · View at Scopus
  16. D. Taylor, F. J. O'Brien, A. Prina-Mello, C. Ryan, P. O'Reilly, and T. C. Lee, “Compression data on bovine bone confirms that a “stressed volume” principle explains the variability of fatigue strength results,” Journal of Biomechanics, vol. 32, no. 11, pp. 1199–1203, 1999. View at Publisher · View at Google Scholar · View at Scopus
  17. M. B. Schaffler, K. Choi, and C. Milgrom, “Aging and matrix microdamage accumulation in human compact bone,” Bone, vol. 17, no. 6, pp. 521–525, 1995. View at Publisher · View at Google Scholar · View at Scopus
  18. X. M. Li, H. F. Liu, X. F. Niu et al., “The use of carbon nanotubes to induce osteogenic differentiation of human adipose-derived MSCs in vitro and ectopic bone formation in vivo,” Biomaterials, vol. 33, no. 19, pp. 4818–4827, 2012. View at Publisher · View at Google Scholar · View at Scopus
  19. D. E. Brehl and T. A. Dow, “Review of vibration-assisted machining,” Precision Engineering, vol. 32, no. 3, pp. 153–172, 2008. View at Publisher · View at Google Scholar · View at Scopus
  20. K. Alam, A. V. Mitrofanov, and V. V. Silberschmidt, “Experimental investigations of forces and torque in conventional and ultrasonically-assisted drilling of cortical bone,” Medical Engineering and Physics, vol. 33, no. 2, pp. 234–239, 2011. View at Publisher · View at Google Scholar · View at Scopus
  21. D. Vashishth, “Rising crack-growth-resistance behavior in cortical bone: implications for toughness measurements,” Journal of Biomechanics, vol. 37, no. 6, pp. 943–946, 2004. View at Publisher · View at Google Scholar · View at Scopus
  22. J. Pujana, A. Rivero, A. Celaya, and L. N. López de Lacalle, “Analysis of ultrasonic-assisted drilling of Ti6Al4V,” International Journal of Machine Tools and Manufacture, vol. 49, no. 6, pp. 500–508, 2009. View at Publisher · View at Google Scholar · View at Scopus
  23. S. S. F. Chang and G. M. Bone, “Thrust force model for vibration-assisted drilling of aluminum 6061-T6,” International Journal of Machine Tools and Manufacture, vol. 49, no. 14, pp. 1070–1076, 2009. View at Publisher · View at Google Scholar · View at Scopus
  24. U. Heisel, J. Wallaschek, R. Eisseler, and C. Potthast, “Ultrasonic deep hole drilling in electrolytic copper ECu 57,” CIRP Annals, vol. 57, no. 1, pp. 53–56, 2008. View at Publisher · View at Google Scholar · View at Scopus
  25. X. Wang, L. J. Wang, and J. P. Tao, “Investigation on thrust in vibration drilling of fiber-reinforced plastics,” Journal of Materials Processing Technology, vol. 148, no. 2, pp. 239–244, 2004. View at Publisher · View at Google Scholar · View at Scopus
  26. B. Azarhoushang and J. Akbari, “Ultrasonic-assisted drilling of Inconel 738-LC,” International Journal of Machine Tools and Manufacture, vol. 47, no. 7-8, pp. 1027–1033, 2007. View at Publisher · View at Google Scholar · View at Scopus
  27. A. Cardoni, A. MacBeath, and M. Lucas, “Methods for reducing cutting temperature in ultrasonic cutting of bone,” Ultrasonics, vol. 44, pp. e37–e42, 2006. View at Publisher · View at Google Scholar · View at Scopus
  28. S. Harder, S. Wolfart, C. Mehl, and M. Kern, “Performance of ultrasonic devices for bone surgery and associated intraosseous temperature development,” The International Journal of Oral & Maxillofacial Implants, vol. 24, no. 3, pp. 484–490, 2009. View at Scopus