Research Article
Fabrication and Permeability Characteristics of Microdialysis Probe Using Chitosan Nanoporous Membrane
Table 2
Cut-off size in diffusion study (data were expressed as mean ± S.D.).
| | 3 kD dextran | 10 kD dextran | 40 kD dextran |
| CG 50 | 12.48 ± 0.9% | 4.53 ± 0.85% | 0.65 ± 0.64% | CG 35 | 12.17 ± 0.84% | 0.92 ± 0.74% | 0.54 ± 0.44% | CG 25 | 13.55 ± 2.2% | 0.95 ± 0.83% | 0.73 ± 0.65% | CG 0 | 0.09 ± 0.01% | 0.08 ± 0.01% | 0.13 ± 0.02% |
|
|