Review Article
Recent Design Development in Molecular Imaging for Breast Cancer Detection Using Nanometer CMOS Based Sensors
Table 1
Read-out noise estimation (worst case).
| Simulate read-out noise | Description | Value | Unit |
| Electronic gain | | 29.6 | electron/ADU | Dark rate | 30.6 ADU/sec electron/ADU | 906 | electron/pixel/sec | Dark current noise | sqrt(dark rate ) | | | Read-out noise | Gain × std_dev(offset)/sqrt(2) | | | | 29.6 * 3.27/1.414 | 69 | electron/pixel (for 500 kHz rate) | SNR | Full well potential/read-out noise | | | | log10(1700000/69) | 87.8 | dB |
|
|