Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents

Table 1

Comparison of three consecutive biometric measurements taken by both ophthalmology residents (Repeatability).

ParameterResident1st measurement
Mean ± SD
(Min–Max)
2nd measurement
Mean ± SD
(Min–Max)
3rd measurement
Mean ± SD
(Min–Max)
ICC (Confidence interval)

(D)1 (40.18–45.61) (40.08–46.30) (40.23–45.67)0.995
(0.993–0.996)
2 (40.08–45.79) (40.13–45.73) (39.94–45.67)0.993
(0.99–0.995)

(D)1 (40.37–47.74) (40.42–47.20) (40.37–47.14)0.993
(0.991–0.995)
2 (40.52–47.07) (40.37–47.20) (40.37–47.27)0.988
(0.984–0.991)

WTW (mm)1 (10.7–13.2) (10.7–13.2) (10.8–13.2)0.937
(0.917–0.954)
2 (10.8–13.2) (10.7–13.2) (10.7–13.2)0.96
(0.947–0.97)

CCT ( )1 (479–677) (482–676) (480–676)0.97
(0.96–0.978)
2 (478–674) (482–673) (481–677)0.96
(0.947–0.971)

ACD (mm)1 (2.8–4.67) (2.81–4.67) (2.8–4.68)0.995
(0.993–0.996)
2 (2.79–4.67) (2.8–4.69) (2.8–4.69)0.995
(0.993–0.996)

PS (mm)1 (3.4–8.3) (3.2–8.4) (3.1–8.3)0.872
(0.832–0.905)
2 (3.3–8.3) (3.2–8.4) (3–8.3)0.905
(0.875–0.929)

AL (mm)1 (20.89–27.23) (20.9–27.2) (20.9–27.23)0.999
(0.998-0.999)
2 (20.89–27.23) (20.91–27.22) (20.91–27.23)0.999
(0.998-0.999)