Research Article

Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents

Table 2

Comparison of three consecutive IOL power measurements taken by both ophthalmology residents (Repeatability).

ParameterResident1st measurement
Mean ± SD
(Min–Max)
2nd measurement
Mean ± SD
(Min–Max)
3rd measurement
Mean ± SD
(Min–Max)
ICC (Confidence interval)

SRK/T (D)1 (10.98–30.86) (11.1–30.9) (10.95–30.93)0.983
(0.977–0.988)
2 (11.53–30.86) (11.22–30.72) (11.02–31)0.996
(0.995–0.997)
Difference (−1.73–0.66) (−0.66–0.6) (−2.85–5.78)

Holladay (D)1 (10.46–31.35) (10.59–31.41) (10.43–31.44)0.983
(0.977–0.987)
2 (11.06–31.35) (10.72–31.21) (10.5–31.53)0.996
(0.994–0.997)
Difference (−1.87–0.63) (−0.71–0.64) (−2.7–6.11)

HofferQ (D)1 (10.58–31.84) (10.7–31.9) (10.53–31.94)0.981
(0.974–0.986)
2 (11.23–31.84) (10.84–31.69) (10.61–32.03)0.995
(0.994–0.996)
Difference (−2.14–0.67) (−0.76–0.68) (−2.74–6.83)

Friedman test.