Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents

Table 4

Comparison of IOL power measurement means obtained by both ophthalmology residents (Reproducibility).

ParameterResidentMean ± SD
(Min–Max)
ICC
(Confidence interval)

SRK/T (D)1 (11.01–30.9) 0.996
(0.995–0.998)
2 (11.26–30.86)
Difference (−0.91–2.01)

Holladay (D)1 (10.49–31.4) 0.996
(0.995–0.997)
2 (10.76–31.36)
Difference (−0.85–2.13)

HofferQ (D)1 (10.6–31.89) 0.996
(0.994–0.997)
2 (10.89–31.85)
Difference (−0.85–2.37)

Paired samples test.