Research Article
Comparison of Two Different Scheimpflug Devices in the Detection of Keratoconus, Regular Astigmatism, and Healthy Corneas
Table 5
Interdevice differences of the two devices in corneas with regular astigmatism.
| | Pentacam | Sirius | values | | Mean ± SD |
| Pachymetry (µm) | | | | Apex | 560.45 ± 36.41 | 590.66 ± 55.62 | <0.0001 | Thinnest point | 554.84 ± 35.50 | 545.88 ± 33.45 | <0.0001 | Keratometry ant. surface (mm) | | | | 3 mm flat meridian | 7.982 ± 0.374 | 8.007 ± 0.378 | 0.654 | 3 mm steep meridian | 7.685 ± 0.344 | 7.631 ± 0.396 | 0.002 | 5 mm flat meridian | 8.023 ± 0.372 | 7.996 ± 0.369 | 0.343 | 5 mm steep meridian | 7.752 ± 0.359 | 7.663 ± 0.385 | 0.002 | 7 mm flat meridian | 8.187 ± 0.430 | 8.042 ± 0.429 | <0.0001 | 7 mm steep meridian | 7.914 ± 0.387 | 7.750 ± 0.387 | <0.0001 | Keratometry back surface (mm) | | | | 3 mm flat meridian | 6.719 ± 0.377 | 6.838 ± 0.419 | 0.001 | 3 mm steep meridian | 6.347 ± 0.363 | 6.317 ± 0.536 | 0.587 |
|
|