Research Article

Comparison of Fourier-Domain and Time-Domain Optical Coherence Tomography in the Measurement of Thinnest Corneal Thickness in Keratoconus

Table 1

Difference and correlation analysis of the TCT from the RTVue-OCT and Visante-OCT.

RTVue-OCTVisante-OCTDifferenceCorrelationICC
MeanSD95% CI-U95% CI-LMeanSD95% CI-U95% CI-LMeanSD(95% LoA)

Normal eyes530.419.7500.2564.5521.518.3493.4555.28.96.9<0.0010.938<0.0010.915 (0.906–0.952)
Keratoconus425.058.2307.7504.6424.455.7310.7499.80.610.20.6040.985<0.0010.984 (0.976–0.990)
Subgroup 1
(keratoconus)
469.526.5433.4518.6466.225.9430.7517.33.27.70.0090.957<0.0010.950 (0.899–0.974)
Subgroup 2
(keratoconus)
376.041.7286.7425.7378.341.3291.3423.0−2.311.80.2230.960<0.0010.959 (0.924–0.978)