OCT Biometry (B-OCT): A New Method for Measuring Ocular Axial Dimensions

Table 2

Interobserver reproducibility results for B-OCT using REVO NX based on the first readings from each session taken by 2 clinicians in a sample of 50 eyes.

Parameter1st observer, mean ± SD2nd observer, mean ± SDSwTRTCoV (%)ICC

AXL (mm)23.93 ± 0.8023.93 ± 0.800.020.040.061.000
CCT (μm)555.72 ± 26.55555.54 ± 27.252.216.120.400.993
ACD (mm)3.58 ± 0.283.59 ± 0.280.060.060.650.993
LT (mm)4.02 ± 0.224.02 ± 0.220.030.070.630.985

AXL: axial length; CCT: central corneal thickness; ACD: anterior chamber depth; LT: lens thickness; Sw: within-subject SD; TRT, test-retest repeatability (2.77 Sw); CoV: within-subject coefficient of variation; ICC: intraclass correlation coefficient.