Research Article
Use of the Charge Transfer Reactions for the Spectrophotometric Determination of Risperidone in Pure and in Dosage Forms
Table 2
Evaluation of accuracy and precision.
| Method | RSP taken, μg/mL | Intra-day accuracy and precision | Inter-day accuracy and precision | RSP found*, μg/mL | % RE | % RSD | RSP found, μg/mL | % RE | % RSD |
| Method A | 5 | 4.98 | 0.43 | 0.41 | 5.03 | −0.49 | 0.23 | 15 | 14.89 | 0.74 | 0.39 | 15.05 | −0.33 | 0.24 | 20 | 19.93 | 0.35 | 0.28 | 20.14 | −0.67 | 0.39 |
| Method B | 5 | 4.98 | 0.42 | 0.79 | 5.11 | −2.28 | 0.79 | 20 | 19.78 | 0.94 | 0.35 | 20.03 | −0.12 | 0.39 | 40 | 39.89 | 0.28 | 0.27 | 40.12 | −0.29 | 0.53 |
|
|
RE: relative error; RSD: relative standard deviation.
*Mean value of 5 determinations.