Abstract

This paper reports the successful fabrication of an impedance-based miniaturized biosensor and its application for ultrasensitive Prostate-Specific Antigen (PSA) detection in standard and real human plasma solution, spiked with different PSA concentrations. The sensor was fabricated using photolithographic techniques, while monoclonal antibodies specific to human PSA were used as primary capture antibodies. Electrochemical impedance spectroscopy (EIS) was employed as a detection technique. The sensor exhibited a detection limit of 1 pg/ml for PSA with minimal nonspecific binding (NSB). This detection limit is an order of magnitude lower than commercial PSA ELISA assays available on the market. The sensor can be easily modified into an array for the detection of other biomolecules of interest, enabling accurate, ultrasensitive, and inexpensive point-of-care sensing technologies.

1. Introduction

Prostate cancer is one of the most widespread cancerous malignancies amongst the male population in the United States [1]. Prostate cancer is the second leading cause (after lung cancer) for all cancer-related deaths amongst males. On average, males in the United States have 1 : 6 odds of being diagnosed with prostate cancer during their lifetime [2]. Currently, there are only two approved prostate cancer screening methods: Digital Rectal Examination (DRE) and Prostate-Specific Antigen (PSA) blood test [3]. DRE, however, possesses low sensitivity and is a function of the skills of the doctor conducting the examination. Additionally, due to hesitation amongst males to undergo the “invasive” exam and variability in the skill levels of the doctors, DRE in many instances leads to the diagnosis of advanced noncurable tumors [4]. Hence, clinicians are increasingly turning to less invasive blood-based diagnostics tests (e.g., serum PSA), which [3] have led to tremendous improvements in early detection and reduced prostate cancer-related mortality and morbidity [5]. The epidemiological importance of prostate cancer therefore warrants a faster, cheaper, and more accurate sensing device for the PSA-based prostate cancer screening and detection.

The generally accepted technique for PSA detection is Enzyme-Linked Immuno Assay (ELISA). However, ELISA and its variations have a detection limit in the high-picomolar ranges. Detection techniques which employ chromatography principle, while offering potentially lower detection limits, are expensive, laborious, and time intensive and are not suited to both routine screening applications and point-of-care diagnostics [6]. As a response to the need for better and sensitive detection methods, numerous novel detection techniques have recently emerged [7]. A proposed alternative to conventional ELISAs is the use of nanoparticles, which has shown promise to achieve lowered detection limits [8, 9]. These methods, however, share some of the drawbacks of chromatography: increased cost, labor, and analysis time. In addition, the inability to reliably estimate a number of nanoparticles attached to one detector biomolecule has made it difficult to quantify the results.

Electrochemical biosensors constitute a promising group of sensing devices that allow increased sensitivities, low cost, low analysis times, affordability, and miniaturized platforms [1017]. There are numerous electrochemical techniques currently being researched towards applications in biosensors, which are described in detail elsewhere [1827]. Among them, Electrochemical Impedance Spectroscopy-(EIS-) based detection is gaining significant interest as a label-free technique for sensitive measurement of target analytes [28]. EIS is a powerful and sensitive technique used to characterize surface-modified electrodes and for the investigation of electrochemical systems and processes [29]. It uses periodic small AC amplitudes and responds to signal change caused by the binding of target analytes to primary antibodies immobilized on the surface of the electrodes [30, 31].

This paper reports successful development of an impedance-based miniaturized biosensor for PSA detection. The biosensor exhibited a detection limit of 1 pg/ml for PSA in human plasma. This sensitivity is an order of magnitude lower than the most sensitive commercial ELISA on the market [32]. Average detection time with the sensor for one sample is around 45 minutes, compared to at least 2.5 hours for a standard ELISA [32]. This research has also addressed and minimized the issue of nonspecific binding (NSB).

2. Materials and Methods

2.1. Chemicals and Reagents

Dithiobis(succinimidyl propionate) (DTSP) and sodium borohydride (NaBH4) were purchased from ThermoFisher Scientific. Purified PSA protein (product # 7820-0604), Monoclonal PSA antibody (anti-PSA, Mab, product # 7820-0217), and monoclonal cortisol antibody (product # 2330-4839) were procured from Abd Serotec. Phosphate buffered saline (PBS) tablets were purchased from Sigma Aldrich. SU-8 50 resist was purchased from Microchem Corp. Blocking buffers (general low-level BB1 and Neptune Block BB2) were obtained from Immunochemistry Technologies, LLC. All other chemicals were of analytical grade and were used without further purification. PBS solution (10 mM, pH 7.4) was prepared by dissolving 1 tablet in 200 ml of deionized water. Working solutions of PSA, anti-PSA, and anticortisol were prepared by dilution in PBS (10 mM, pH 7.4).

2.2. PSA Sample Collection and Storage

Human serum and plasma samples, collected at Moffitt Cancer Center and participating clinics, were used in this study. The study was approved by an Institutional Review Board (IRB), and informed consent was obtained from all participants prior to a blood draw. The deidentified samples were obtained from African-American males with clinically diagnosed Prostate cancer (cases), and with no clinical evidence of Prostate cancer (controls). Samples were labeled, aliquoted, and stored at deep freezer (−80°C) until further usage according to the Moffitt Cancer Center guidelines [33]. None of the samples have undergone more than 2 freeze-thaw cycles. All samples were tested for PSA in triplicates using commercial ELISA kit and the values were recorded.

2.3. Measurement and Apparatus

Electrochemical impedance spectroscopy (EIS) was utilized to characterize the EA/Anti-PSA/DTSP/IDμE bioelectrode and to estimate PSA concentration. EIS measurements were carried out at equilibrium potential without external biasing in the frequency range of 0.5–105 Hz with a 5 mV amplitude using Autolab Potentiostat/Galvanostat (Eco Chemie, Netherlands). EIS measurements were carried out using 65 μl of PBS solution (10 mM, pH 7.4) containing a mixture of 5 mM (Ferrocyanide) and 5 mM of (Ferricyanide) that is 5 mM as a redox probe. In present studies, Nyquist plots have been utilized to study the change in charge transfer resistance ( ) at sensor-solution interfaces with changing concentration of PSA.

2.4. Test Chip Fabrication

The Biosensor test chips were fabricated on an oxidized 4′′ silicon wafer using standard photolithography techniques, as described previously [34, 35]. Briefly, Cr/Au (200/2000 Å) layers were deposited using e-beam evaporation and were patterned through lift-off (Figure 1(a)). IDμE with 5 μm wide electrode fingers and with a pitch of 10 μm were used in this work. As a final step, an SU8 chamber was patterned around the electrodes using the SU8 50 resist to create a sample well around these electrodes, and hard baked at 200°C to improve its resistance against hard solvents like acetone. Figure 1(b) shows the actual photograph of a Biosensor reagent chamber under optical microscope (10x).

2.5. Self-Assembled Monolayer (SAM) Preparation and Antibody Immobilization

The process was conducted as described previously [35, 36]. Briefly, the Biosensor chips were precleaned with acetone, isopropyl alcohol, and de-ionized water, and exposed to 2 mg/ml solution of DTSP in acetone for 1 hr for SAM formation. DTSP solution was first reduced using NaBH4 and then dispensed on the precleaned chips at room temperature. The DTSP SAM modified electrodes were then rinsed with acetone to remove any unbound DTSP followed by rinsing in water. The electrodes were then utilized for antibody immobilization. PSA antibodies were covalently attached to DTSP self-assembled monolayer by incubating the electrode in 65 μl of 1 μg/ml antibody in PBS solution (10 mM, pH 7.4) for 1 hr. Covalent binding (amide bond formation) results from the reaction between amino group of antibody and reactive succinimidyl group of the DTSP on the SAM surface. The sensor (Anti-PSA/DTSP/IDμE) was washed thoroughly with PBS (10 mM, pH 7.4) to remove any unbound biomolecules followed by a 10-minute washing with ethanolamine (EA) (1%). EA was used to block any unreacted succinimidyl groups on DTSP SAM and to remove extra unbound antibodies onto the electrode surface. Figure 2 schematically illustrates (a) IDμE chip, (b) step-by-step immunochemical reaction on the electrode surfaces, and (c) proposed detection system. The fabricated detection limits were characterized using electrochemical impedance technique. Each sample was run in triplicates to ensure reproducibility, and an average result of three runs was used.

It is noteworthy, that due to the small size of the IDμE chip, any change in the sample volume or in the nature of the solutions or dielectric properties of material may affect the results of sensing. Therefore, in the present work, all experiments were carried out under identical conditions of solution volume and materials, and change in charge transfer resistance (difference of before and after external incubation of PSA concentration) for PBS containing (measurement buffer) was measured for PSA estimation. As all the conditions were kept identical during EIS measurement and change in signal was used for estimation, the factors such as solution volume or material get cancelled. Hence, the change in the EIS signal is attributed solely to changing PSA concentration which on interaction with surface bound antibody results in formation of the insulating layer on the surface, thus causing increase of charge transfer resistance for measuring buffer.

3. Results and Discussion

3.1. Biosensor Testing with the PSA Solutions in PBS

Biosensor was first utilized to detect PSA molecules in PBS in the concentration range 1 pg/ml–10 ng/ml (Figure 3(a)). This range was chosen empirically, based on the interest for meeting both current needs in PSA sensitivity (usually up to 10 ng/ml in clinics) and establishing the lower detection limit. For each concentration, the EA/Anti-PSA/DTSP/IDμE bioelectrode was incubated in PSA solution for 30 minutes, followed by PBS washing and EIS spectra recording using PBS (10 mM, pH 7.4) containing 5 mM as a redox probe. From Figure 3(a), it is clear that (diameter of the Nyquist plots) increased with increasing PSA concentration. The increase in is attributed to the binding of PSA to immobilized anti-PSA on EA/Anti-PSA/DTSP/IDμE bioelectrode, producing a barrier layer that inhibits the charge transfer for a redox probe, as relative change in EIS data has more significant information than absolute value for sensing applications. The change in was used for calibration. A plot of the change in values and the logarithm of PSA concentrations reveal a linear detection range for PSA concentrations in the range of 1 pg/ml to 10 ng/ml (data not shown). A calibration curve for normalized data is shown on the Figure 3(b).

Due to lack of industrial level controls during microfabrication and chemical activity, variations in impedance of individual electrodes and antibody modified electrodes were observed. Therefore, to confirm that the observed change in impedance was due to surface modification and not due to superimposed effects, the data was normalized to charge transfer resistance for desired concentration [ ]/[charge transfer resistance of blank EA/Anti-PSA/DTSP/IDμE bioelectrode ]. In Figure 3(b), plot of versus the logarithm of cortisol concentration shows the results of triplicate set. After normalization, all electrodes with different impedances for detection limits with attached antibodies exhibited similar response within the 4% error for each concentration. The normalized data curve (Figure 3(b)) can be characterized using a linear relation; . It reveals the linear range of 1 pg/ml to 10 ng/ml with the detection limit of 1 pg/ml and correlation coefficient of 0.959. Further, to account for the variation in initial impedance values of individual electrodes and to avoid superimposed effects of multielectrode measurement, all experiments were carried out using a step-by-step approach to increase PSA concentration. Similar step-by-step concentration studies have been reported by other researchers and help avoid superimposed effects of multielectrode measurement [3742].

As is seen from Figure 3(b), the correlation coefficient for the calibration curve is 0.959, suggesting a relatively weak linear trend. This effect can be attributed to the interfering effects of Nonspecific binding (NSB) which is a common obstacle in all bioassays, employing immunological affinity principles. In this study, general BSA-based (BB1) and increased strength nonmammalian proteins-based (BB2) blocking buffers and 30 to 60 min blocking times were studied. However, no significant change in NSB interference was observed with the introduction of different blocking techniques or extra strength blocking buffers (data not shown). To handle NSB issues in a better way, new experiments with diluted serum as blocking were designed and performed.

3.2. Human Plasma Testing with Biosensor

It was hypothesized that a human plasma sample, even when substantially diluted, will have enough proteins to block the initial electrode and can minimize subsequent NSB during bioassays. Hence, in order to address the NSB issues, a real plasma sample with very low PSA concentration of 0.4 ng/ml was diluted 1000 times with PBS. This solution was then used as a diluent for the rest of the experiments. The level of PSA in the diluents was an order of magnitude lower than the concentration of PSA in samples, thus ensuring that diluent PSA does not interfere with sample PSA detection. It was observed that the use of diluent and blocker masked major NSB issues and established that sensors can be used with human serum samples without a loss in sensitivity and selectivity. To perform the experiments, diluent was subsequently spiked with different PSA concentrations. Based on previous results (Figure 3), 1; 10 and 100 pg/ml of PSA were chosen as initial testing points. Knowing that PSA concentration in the diluent was 0.4 pg/ml, the actual PSA values in the samples were 0.4, 1.4, 10.4, and 100.4; pg/ml.

The testing was carried out in the same low to high concentration order as described earlier for standard PSA solutions. The results shown in Figure 4 reveal a clear dependence of the impedance resistance with increasing PSA concentration, indicating PSA binding. Each step was resumed after PBS washing to ensure complete removal of unbound molecules.

It is important to note that due to a pilot nature of this work, no formal statistical analysis was performed. The overall testing procedures with Biosensor were as follows: (1) SAM formation on gold electrode; (2) anti-PSA incubation + wash + impedance reading; (3) base solution (“blocker”, 0.4 pg/ml PSA) incubation + wash + reading; (4) 1 pg/ml incubation + wash + reading; (5) 10 pg/ml incubation + wash + reading; (6) 100 pg/ml incubation + wash + reading.

Figure 5 presents the calibration curve for data obtained from Figure 4 where before first run bioelectrode was treated and blocked with diluent and it is clear from Figure 5 that there is a strong linear concentration dependence on the impedance increase in resistance with increased PSA concentration. The correlation coefficient for this experiment was found to be 0.995, indicating strong linear dependence. The normalized data curve (Figure 5) can be characterized using .

3.3. Selectivity Studies

Selectivity is a major and well-known pitfall in biosensing techniques. To ensure the selectivity of PSA binding, increasing concentrations of cortisol antibody solutions in PBS were tested on the electrodes functionalized with captured anti-PSA. Anti-cortisol was selected as a random protein which can contribute to nonselective binding with 0; 1; 10; 100 and 1000 pg/ml solutions of anti-cortisol in PBS. The testing was done in identical conditions. Ideally, no increase in resistance should be observed if NSB is completely absent. The results shown in Figure 6 reveal a slight increase in resistance with increasing anti-cortisol concentrations. However, the observed change is very low as compared to PSA (15.5% highest increase compared to 312% increase for lowest PSA concentration), indicating that NSB is in fact present although negligibly small.

4. Conclusions

It has been shown that the EA/Anti-PSA/DTSP/IDμE bioelectrode-based, impedimetric electrochemical immunosensor, can be used for ultrasensitive and selective PSA detection. EA/Anti-PSA/DTSP/IDμE bioelectrode exhibits linear behavior in the concentration range of 1 pg/ml to 100 pg/ml with detection limit of 1 pg/ml. This detection limit is an order of magnitude lower than the most sensitive commercially available PSA ELISA kit on the market. The average analysis time of 45 minutes is shorter than commercial ELISA (at least 2.5 hrs). The usage of diluted plasma sample as a diluent and blocker allowed masking of major NSB issues. The sensor was found to be selective against cortisol antibody in the concentrations 1–1000 pg/ml and was tested as a proof of concept with the human plasma sample, spiked with increasing PSA concentrations. The same principle could be applied for the detection of other biomolecules of interest.

Acknowledgments

This work was partially supported through NSF Award 0700659 Nanoengineered, Manufacturable, Ion-Implantation Seeded Silica Nanowires for Sensitive BioScreening, UF—Moffitt Pilot Award Prostate Cancer Biomarker Study in African American Men; BiTT Award: An Automated Cell Health Monitoring System (CHMS) Based on Electrical Impedance; the National Institute of Health—NIH 1 P20 MD003375-01.