Abstract

The field measurement of liquid viscosities, especially the high viscous liquids, is challenging and often requires expensive equipment, long processing time, and lots of reagent. We use quartz crystal microbalances (QCMs) operating in solution which are also sensitive to the viscosity and density of the contacting solution. QCMs are typically investigated for sensor applications in which one surface of QCM completely immersed in Newtonian liquid, but the viscous damping in liquids would cause not only large frequency shifts but also large losses in the quality factor leading to instability and even cessation of oscillation. A novel mass-sensitivity-based method for field measurement of liquid viscosities using a QCM is demonstrated in this paper and a model describing the influence of the liquid properties on the oscillation frequency is established as well. Two groups of verified experiments were performed and the experimental results show that the presented method is effective and possesses potential applications.

1. Introduction

The measurement of liquid viscosities is one of the most important steps in a variety of research and industrial fields. There are many techniques including capillary [1], falling body [2], oscillating [3], and ultrasonic [4] methods that can measure the viscosity of liquids. However, these commercial devices are usually costly, time consuming, and limited in the measurement range.

During the last few decades, quartz crystal microbalances (QCMs) are increasingly being studied as a sensitive, rapid, high-performance, and inexpensive microsensors capable of performing in liquid environments [512]. A QCM is a shear mode device consisting of a thin quartz disk sandwiched between two circular, metallic electrodes of the same diameters. The electrodes are used to excite the resonator whose frequency is inversely proportional to the thickness of the quartz plate. As a highly sensitive mass sensor, QCMs have been used in various areas of science and technology, such as mass detectors in vapors or solutions [1316], biosensors [17, 18], and electrochemical microbalances or nanobalances [1921].

The linear relationship between the mass change and the resonance frequency shift of the QCM is described by the Sauerbrey equation [22]:where and are mass change and frequency shift, respectively; is the operating frequency of the QCM; is the active area of the QCM electrodes; and and are the density and shear modulus of the piezoelectric quartz crystal, respectively.

It should be noted here that the Sauerbrey equation is established on the assumption that the mass change attached on the QCM surface is a rigid and even thin film. So the linear relation between the adsorbed mass and the change in frequency is not necessarily valid for viscoelastic films. Therefore, the Sauerbrey equation is invalid for the measurement of liquid parameters.

Kanazawa and Gordon II offered an additional method to measure Newtonian liquid using a QCM, namely, the total immersion of one surface of the QCM in sample liquid, and determined the behavior of the crystal/fluid system by examining the coupling of the elastic shear waves in the crystal to the viscous shear weaves in the sample liquid. In this manner, the resonance condition derives directly from the matching of appropriate boundary conditions on the shear waves. Thus, the Kanazawa model that reveals the relative relationship between the frequency shift of the QCM and the density and viscosity of the sample liquid has been developed [23]:where is the operating frequency of the QCM; and are the density and viscosity of the sample liquid, respectively; and are the density and shear modulus of the quartz having the values = 2.648 × 103 kg/m3 and = 2.947 × 1011 g/cm·s2, respectively.

However, there are still some inevitable drawbacks when using Kanazawa model for on-field measurement of liquid properties; that is, this method is difficult to be applied and spread widely in on-field testing because it is sample wasteful, requires complex instruments, and is difficult to operate. In addition, immersing the QCM in liquid will damp the amplitude, and the damping becomes seriously detrimental when immersed in highly viscous liquid, which will make the measurements hard or even causes the failure of oscillation.

This new mass-sensitivity-based approach presented in this paper could promote the application of QCMs in on-field testing of liquid viscosities and enable the measurement of highly viscous liquids. Thanks to the advantages of this new approach including simplicity of operation, saving reagent and time, real-time output, and label-free analysis, this novel approach will achieve a considerable potential application prospect in on-field measurements of liquids.

2. Proposed Method

The frequency shift caused by a localized or nonuniform mass attached on the QCM electrodes is given by [24]where is the mass-sensitivity function, in hertz per kilogram; is the effective added mass; is the radius for the mass deposit on the electrode; and are the polar coordinates of the point at which the mass is added.

The research of Josse et al. shows that the mass-sensitivity distribution of QCM devices can be represented by the following equation [25]:where is the Sauerbrey sensitivity constant, with a value of 1.78 × 1011 Hz·cm2/kg; is the particle displacement amplitude function; and is the distance from the center.

The particle displacement amplitude function is solution of the following particle displacement amplitude equation [26]:where represents the wavenumber; is a constant determined by the elastic stiffness constants and the piezoelectric constants of crystal quartz. Since the particle displacement and shear strain field are continuous at (where is the radius of electrode), according to these boundary conditions, the boundary equations of QCM with “m-m” type electrode are obtained as [27]

On the basis of these boundary conditions, can be determined. And the mass-sensitivity function of QCM with “m-m” type electrode is then determined [28].

Take two QCMs used in this study (AT-cut 10 MHz fundamental QCM and AT-cut 10 MHz 3rd overtone QCM), for example. The diameters of crystal and gold-electrode are 8.7 mm and 4 mm, respectively. The thickness of the gold-electrode is 1000 Å. The profile of mass-sensitivity distribution can be obtained as shown in Figures 1 and 2, respectively.

As expected, the mass-sensitivity distribution of the two types of QCMs is an approximate Gaussian curve. The highest sensitivity point appears at the center of electrode () and then decreases exponentially as the distance from the center () increases. So, within the electrode area, the mass sensitivity of the QCM can be described as a Gaussian function, like the following:where is the highest value of sensitivity at the center of the electrode (for AT-cut 10 MHz fundamental QCM = 6.23 × 1012 and for AT-cut 10 MHz 3rd overtone QCM = 2.02 × 1012); is the radius of the gold-electrode; and is a constant, for the two types of QCMs having the values 2.298 and 0.797, respectively.

As Figure 3 shows, when a small single-drop of liquid (in microliter level) loaded on the center of the electrode, the radius of single-drop much less than the electrode radius and their centers coincided; we can assume that is changed a little within the range of . As can be seen from Figures 1 and 2, is relatively smooth near the center of the electrode. In addition, the place loaded by liquid which has the highest mass sensitivity will contribute to lowering of the limit of detection.

The research of Hillier and Ward shows that the effective mass is related to the liquid decay length when the QCM is covered by a viscoelastic film [24]: where and are the density and viscosity of the liquid, respectively.

Based on the analysis above, as shown in Figure 4, the effective mass could be divided into (the effective mass of the liquid droplet contained within decay length for base radius from center to ) and (the effective mass of the liquid droplet contained within decay length for base radius from to ), and the frequency shift of QCM can be calculated aswhere , , and and are the frequency shift caused by and , respectively.

In general, the decay length of liquid is very small (in micrometer level), so is also in micrometer level; that is, is an infinitesimal. It is obvious that is an infinitesimal of higher order than . As a consequence, is very small and could be ignored. So can be calculated as where .

Equation (11) could be obtained through integration: where .

Since , must be very small; then (12) is tenable: So (11) could be simplified to

In summary, considering that the density of liquids is usually known or can be measured easily and accurately using conventional methods, so the viscosity of liquids could be quickly and conveniently obtained by measuring the frequency shift of QCM and the radius of liquid drop according to (13).

3. Experimental Results

3.1. Experimental Setup

Figure 5 shows a schematic of the overall experimental setup. At room temperature (about 20°C), a single-drop of sample liquid was loaded onto the center of the QCM electrode using a microinjector. Then the response of QCM was measured with a vector network analyzer (Agilent E5062A). And the diameter of the single-drop was measured with a tool-microscope (TM-505). The microinjector was purchased from Shanghai Gaoge Industrial and Trading Co., Ltd. (Shanghai, China). The tool-microscope was purchased from Shanghai BM Optical Instruments and Manufacture Co., Ltd. (Shanghai, China).

It is worth adding that a high accuracy microscope was used to acquire high testing precision during experiments. But in practical on-field measurements, sophisticated high microscope is not needed; an ordinary magnifier loupe with scale can meet the measure requirements (with the magnifying power of about 100). Likewise, the expensive, sophisticated, and cumbersome vector network analyzer is also not needed and could be replaced by a small frequency meter. The computer used in these experiments is not necessary as well in practical on-field measurements; these simple calculations could be done by an ordinary calculator.

3.2. The Measurement of Low Viscosity Liquid (Take Pure Water as an Example)

Two types of QCMs were used in this study: AT-cut 10 MHz fundamental QCM and AT-cut 10 MHz 3rd overtone QCM. The only difference between them is the thickness of the crystal. The thickness of crystal of them is about 0.167 mm and 0.5 mm, respectively. The thickness of the gold-electrode of them is 1000 Å. The diameter of crystal and gold-electrode is 8.7 mm and 4 mm, respectively. They were purchased from Tongfang Guoxin Electronics Co., Ltd. (Hebei, China).

At room temperature (about 20°C), about 0.5 μL of pure water ( = 1.0022 mpa·s and = 0.9982 × 103 kg/m3) was loaded onto the center of QCM electrode using a microinjector. The response of AT-cut 10 MHz fundamental QCM before and after loading is shown in Figures 6 and 7, respectively. The response of AT-cut 10 MHz 3rd overtone QCM before and after loading is shown in Figures 8 and 9, respectively. The radius of the single-drop of water loaded on the two QCMs was measured as 0.57 mm and 0.56 mm using tool-microscope, respectively.

As can be seen from Figures 6 and 7, the resonant frequency of AT-cut 10 MHz fundamental QCM declined from 9.972460 MHz to 9.971873 MHz. So we can calculate the frequency shift of QCM = 587 Hz. Figures 8 and 9 showed that the resonant frequency of AT-cut 10 MHz 3rd overtone QCM declined from 10.008099 MHz to 10.007919 MHz. So the frequency shift caused by the single-drop of water is 180 Hz ( = 180 Hz).

According to (13), we can figure out the theoretical frequency shift of each QCM: for AT-cut 10 MHz fundamental QCM and AT-cut 10 MHz 3rd overtone QCM, the theoretical frequency shift is 620 Hz and 186 Hz, respectively. So the errors between theoretical values and experimental values are 5.6% and 3.3%, respectively, and show that the experimental results agree well with the theory.

3.3. The Measurement of High Viscosity Liquid (Take Organic Reagent as an Example)

Other two types of QCMs were used to test the organic reagent: QCM-4.5 and QCM-3; the only difference between them is the diameters of the gold-electrodes, which are 4.5 mm and 3 mm, respectively. The electrode thicknesses of them are all about 1000 Å. The thicknesses and diameters of the quartz crystal wafers (with fundamental operating frequency of 10 MHz) are all about 0.167 mm and 8.7 mm, respectively. They were purchased from Tongfang Guoxin Electronics Co., Ltd. (Hebei, China). The organic reagents were purchased from BP Corporate Introduction, and the density and viscosity of it are ρ = 0.842 × 103 kg/m3 and η = 110 mpa·s at room temperature, respectively.

As shown in Figure 5, at room temperature (about 20°C), a drop of organic reagent was loaded near the center of the electrode of QCM using a microinjector; then the response of QCM and the diameter of single-drop were measured and recorded. Ten experiments were carried out with different sizes of droplets using two types of QCMs (five experiments for each type of QCM, and each experiment was carried out using a new QCM).

Take the second experiment (using a QCM-4.5) and sixth experiment (using a QCM-3), for example. In the second experiment, the responses of the QCM-4.5 without and with the single-drop loading which has a radius of 0.94 mm are shown in Figure 10. The resonant frequency of the QCM-4.5 declined from 9992915 Hz to 9987602 Hz, so the frequency shift can be calculated as = 5313 Hz. In the sixth experiment, the response of the QCM-3 without and with the single-drop loading which has a radius of 0.42 mm is shown in Figure 11. The resonant frequency of the QCM-3 declined from 10004584 Hz to 10003180 Hz, so the frequency shift can be calculated as = 1404 Hz.

The results of all ten experiments are as shown in Table 1, where is the radius of droplets and and are the experimental and theoretical frequency shift, respectively. Moreover, the errors between experimental and theoretical values were calculated according to . The QCM used in the first five experiments is QCM-4.5 and QCM used in the last five experiments is QCM-3. The comparisons between experimental and theoretical values are illustrated in Figure 12 (using QCM-4.5) and Figure 13 (using QCM-3) according to experiment sequence number as abscissa and frequency shift (in Hz) as coordinate.

As can be seen from Table 1 and Figures 12 and 13, to both types of QCMs, the absolute values of errors between theoretical and experimental frequency shift are all less than 8.2%, so the experimental data can be thought of in accordance with the theoretical inferences.

3.4. Error Analysis

There are several reasons that account for the errors between experimental and theoretical values.(a)Operation errors: it is inevitable to have operation error in the process of experiments, such as the fact that the droplet cannot be loaded on the absolute center of the electrode accurately.(b)Measurement errors: both of systematic errors of tool-microscope and visual perception errors of human will cause measurement error during the measurements of droplet radius with tool-microscope.(c)Theoretic errors: in the above process of theoretic deriving, approximate treatments are made in (7), (11), and (13) to simplify the calculations. By the verification of error analysis, it should be noted that the errors caused by approximate treatments in theory are less than 3%, and the errors will decrease with the reduction of the droplet radius.(d)Disturbing errors: disturbance of environments such as humidity, temperature, and airflow will cause disturbing errors inevitably.(e)In addition, the vibration mode of quartz crystal is likely to be affected by the load distribution, especially the asymmetric or heterogeneous load distribution, and then impact the sensitivity distribution. And this issue is worthy of further research.

If the above problems can be solved, the errors between experimental and theoretical values could be controlled at an ideal level. And it leads to the direction of future study of this problem.

4. Conclusions

A novel mass-sensitivity-based method to measure liquid viscosities using a QCM is introduced in this paper, and the relative relationship between the frequency shift and the viscosity of sample liquid is also revealed. Lots of scientific studies and practical applications need to measure liquid viscosities, and this new method has broad prospects for application in on-field testing due to its advantages including reagent-saving, wide response spectrum, simplicity of operation, real-time output, and label-free analysis. At the same time, it could enhance the application of QCMs in various research and industrial fields.

Conflict of Interests

The authors declare that there is no conflict of interests regarding the publication of this paper.