Possibilities of High Resolution Inductively Coupled Plasma Optical Emission Spectrometry in the Determination of Trace Elements in Environmental Materials
Table 1
Specification of ICP-OES spectrometer: HORIBA Jobin-Yvon ULTIMA 2 (France) and ICP operating conditions.
Monochromator
HORIBA Jobin-Yvon ULTIMA 2
Mounting
Czerny-Turner, focal length 1 m
Grating
Holographic, 2400 grooves mm−1
Wavelength range
first and second order
Entrance slit
0.015/0.02 mm
Exit slit
0.02/0.08 mm
Practical spectral bandwidth
5 pm in the 2nd order from 160 nm to 320 nm; and 10 pm in the 1st order from 320 to 800 nm
Detectors
High dynamic detectors based on PMT’s
Radio frequency generator
Solid state RF 40.68 MHz
Frequency power output
40.68 MHz 0. 5–1.55 kW
Nebulizer
Meinhard, concentric glass, type TR 50 C1
Spray chamber
JY Glass cyclonic spray chamber
Plasma torch
Fully demountable torch
Pump
Peristaltic, two channels, twelve-roller
Operating conditions
Incident power, (kW)
1.00
Outer argon flow rate, (L min−1)
13
Nebulizer, pressure, (bar)
3.2
Carrier gas flow rate, (L min−1)
0.4
Sheath gas flow rate, (L min−1)
0.2
Sum = carrier gas + sheath gas flow rate, (L min−1)