Possibilities of High Resolution Inductively Coupled Plasma Optical Emission Spectrometry in the Determination of Trace Elements in Environmental Materials

Table 6

Determination of Zn, Cr, Pb, and Cu in soil sample, obtained by ULTIMA 2: mean values [ ] for ( replicates) and the confidence interval of the mean value [ ] for statistical confidence % and (column 2); relative standard deviation (RSD %) (column 3). Columns 5 and 6 list the corresponding mean values [ ] for statistical confidence % and (column 5) and RSD, %, obtained by D.C. Arc-OES.

ULTIMA 2, ICP-OESD.C. Arc-OES
Selected prominent line, wavelength, λ, nmConcentration, (µg g−1)RSD, %Selected prominent line, wavelength, λ, nmConcentration, (µg g−1)RSD, %

Zn II 202.54877.5 ± 1.21.6Zn I 328.23375 ± 4.25.6
Cr II 267.71654.0 ± 1.01.9Cr II 267.71655.0 ± 3.05.5
Pb II 220.35322.0 ± 0.52.3Pb I 283.56322 ± 1.04.6
Cu I 324.75415.0 ± 0.42.7Cu I 324.75412 ± 0.65.0