Research Article
Determination of As, Cd, and Pb in Tap Water and Bottled Water Samples by Using Optimized GFAAS System with Pd-Mg and Ni as Matrix Modifiers
Table 2
Temperature program used in the determination of As.
| | Temperature, °C | Ramp, s (N2, L/min) | Hold, s (N2, L/min) | As | Cd | Pb | As | Cd | Pb |
| Drying | 140 | 130 | 120 | 15 (0.2) | 25 (0.2) | 20 (0.2) | 15 (0.2) | 10 (0.2) | 10 (0.2) | Ashing | 600 | 280 | 380 | 10 (0.2) | 10 (0.2) | 10 (0.2) | 10 (0.2), 1.5 (0) | 5.0 (0.2), 1.0 (0) | 10.0 (0.2), 1.0 (0) | Atomization | 2300 | 1800 | 2000 | 0.9 (0) | 0.8 (0) | 0.9 (0) | 1.0 (0) | — | — | Cleaning | 2600 | 2500 | 2500 | 1.0 (1.0) | 2.0 (1.0) | 2.0 (1.0) | 3.0 (1.0) | — | — |
|
|