Research Article

Portable Gas Analyzer Based on Fourier Transform Infrared Spectrometer for Patrolling and Examining Gas Exhaust

Table 4

Maximum testing error of relative error.

ItemConcentration rangeMaximum error or relative error

1≤10 × 10−69.3 × 10−6
210~100 × 10−619.9 × 10−6
3100~1000 × 10−69.5%
41000~10000 × 10−64.7%
510000~100000 × 10−63.5%
6≥1000002.2%