Assessment of Coverage and Quality of Selected Clinical Chemistry Tests among Medical Laboratories of Health Facilities in Jimma Zone, South West Ethiopia

Table 3

Normal range, stated value, target value (manufacturer), mean of participating laboratories, and acceptable values based on TEa ± 20%, ±10%, ±15%, and ±20 for each analytes ALT, glucose, creatinine, and total bilirubin, respectively,from Humatrol P control sample given in IU/L and mg/dl.

Type of test Normal rangeStated value(author) Target value (manufacturer) Mean of participating laboratories Acceptable range based on stated valueAcceptable range based on mean value

ALT112.0-180.0139.72.4-142.0110.1118.7-167.688.1-132.2

Glucose229.0-317.0292.515.2-273.0218.4263.3-321.8196.6-240.2

Creatinine3.6-5.64.24.2-4.84.23.6-4.93.6-4.8

Bilirubin total3.4-5.86.972.6-4.34.45.5-8.23.5-5.3