Research Article

Machine Vision Based Automatic Detection Method of Indicating Values of a Pointer Gauge

Table 4

Comparisons between manual reading and visual detection.

The pressure output of the checkpoint (MPa)The indicating value of visual detectionThe indicating value of manual reading Comparison (error 1 − error 2)
Positive stroke Error 1:
(MPa)
Relative error 1: (%)Positive stroke Error 2:
(MPa)
Relative error 2: (%)
Standard gauge: (MPa)Tested gauge: (MPa)Standard gauge:
(MPa)
Tested gauge:
(MPa)

0000000000
0.0250.0243430.0256500.0013105.370000.0240.0260.0028.3330.08202
0.0500.0487270.0502000.0014733.022960.0490.0510.0024.0820.03533
0.0750.0743870.0754700.0010801.456440.0750.0760.0011.3330.01225
0.1000.1004080.1001000.0003080.306750.0990.1010.0022.020.02051
0.1250.1250600.1260000.0009400.751640.1250.1260.0010.80.00706
0.1500.1497620.1509000.0011400.759870.150.1520.0021.3330.01219
0.1750.1746130.1776260.0030101.725530.1750.1730.0021.14290.01440
0.2000.1983160.2022630.0039501.990260.20.2030.0031.50.01105
0.2250.2250530.2261460.0010900.485660.2240.2270.0031.3390.01230
0.2500.2493440.2490570.0002870.115100.2490.2520.0031.2050.01233