Research Article

Applying an Open-Source Segmentation Algorithm to Different OCT Devices in Multiple Sclerosis Patients and Healthy Controls: Implications for Clinical Trials

Table 4

Comparison of the rates of change of retinal layer thickness between Spectralis and Cirrus HD-OCT.

LayerRate of change of layer thickness, Spectralis
mean (95% CI) (μm/year)
valueRate of change of layer thickness, Cirrus
mean (95% CI) (μm/year)
value

mRNFL−0.28 (−0.56, −0.001)0.049−0.09 (−0.28, 0.09)0.326
GCIP−0.59 (−0.89, −0.27)<0.001−0.66 (−0.96, −0.36)<0.001
INL−0.07 (−0.16, 0.01)0.102−0.06 (−0.16, 0.03)0.198
INL + OPL−0.08 (−0.27, 0.11)0.419−0.11 (−0.26, 0.04)0.157
ONL−0.06 (−0.25, 0.13)0.526−0.11 (−0.27, 0.05)0.196
ONL + PR0.05 (−0.21, 0.31)0.689−0.16 (−0.39, 0.07)0.166
RPE0.004 (−0.12, 0.12)0.947−0.002 (−0.13, 0.12)0.971

mRNFL: macular retinal nerve fiber layer; GCIP: ganglion cell layer + inner plexiform layer; INL: inner nuclear layer; INL + OPL: inner nuclear layer + outer plexiform layer; ONL: outer nuclear layer; ONL + PR: outer nuclear layer + photoreceptor segments; RPE: retinal pigment epithelium.